Data Quality Reports for Session: 185157 User: narenathreyask1 Completed: 04/21/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D140822.13NIM/MERGESONDE/M1 - Suspect rh_scaled valuesnimmergesonde2maceM1.c1Suspect


DQRID : D140822.13
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/10/2006000001/07/20072359Suspect
Subject:
NIM/MERGESONDE/M1 - Suspect rh_scaled values
DataStreams:nimmergesonde2maceM1.c1
Description:
MERGESONDE sometimes uses bad or questionable precipitable water vapor data to scale the 
relative humidity profiles provided by the rh_scaled field. In addition, this bad input 
data is not properly indicated by the vapor_source or qc_rh_scaled fields. This problem 
appears to typically occur when there is moisture impacting the microwave radiometer causing 
large spikes in brightness temperatures. During these periods the rh_scaled values are 
often saturated at 101% for much of the profile. 

A new version of the MERGESONDE VAP has corrected this problem, and these data will 
eventually be corrected by a reprocessing task.
Measurements:nimmergesonde2maceM1.c1:
  • vapor_source
  • rh_scaled


Back To Table of Contents