DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D161025.6 | ENA/AOS/C1 - Instrument mentors working at the site | enaaospsap3wC1.01 | Suspect |
D170303.12 | ENA/AOS/C1 - Dry air supply intermittently not working | enaaospsap3wC1.01 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/26/2016 | 0830 | 10/01/2016 | 0100 | Suspect |
Subject: | ENA/AOS/C1 - Instrument mentors working at the site |
DataStreams: | enaaospsap3wC1.01 |
Description: | Instrument mentors working at the site. Sample lines and electrical connections being reconfigured. Instruments that had no work done on them may still be affected. |
Measurements: | enaaospsap3wC1.01:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/20/2016 | 0000 | 03/03/2017 | 0000 | Does not affect quality |
Subject: | ENA/AOS/C1 - Dry air supply intermittently not working |
DataStreams: | enaaospsap3wC1.01 |
Description: | Pentras dry air supply intermittently not working. Dry air missing for the humidigraph and the HT-DMA. Dry air supply for those instruments partially replaced by regular undried compressed air, lowering the delivery pressure. No direct effect on other instruments seen. |
Measurements: | enaaospsap3wC1.01:
|