DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140609.1 | ENA/MWR3C/C1 - Reprocessing Notice (RID 1134) | enamwr3cC1.b1 | Does not affect quality |
D140625.3 | ENA/MWRP/C1 - Bad calibration | enamwrpC1.b1 | Incorrect |
D150219.2 | ENA/MWRP/C1 - Reprocess: Calibration drift | enamwrpC1.b1 | Incorrect |
D150303.6 | ENA/MWRP/C1 - Occasional outages of surface T/RH readings | enamwrpC1.b1 | Suspect |
D160427.7 | ENA/C1 - Reprocess: Slightly incorrect location fields | enamwr3cC1.b1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/24/2014 | 0000 | 06/11/2014 | 1912 | Does not affect quality |
Subject: | ENA/MWR3C/C1 - Reprocessing Notice (RID 1134) |
DataStreams: | enamwr3cC1.b1 |
Description: | See below for details of reprocessing that was completed on 20140902. Original DQR category: 4 (Red, Incorrect). ----------------------------------------------------------- Occasionally the brightness temperatures drop to zero. The problem was identified with occasional missing lines in the raw data. |
Measurements: | enamwr3cC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/19/2014 | 0000 | 06/04/2014 | 1459 | Incorrect |
Subject: | ENA/MWRP/C1 - Bad calibration |
DataStreams: | enamwrpC1.b1 |
Description: | The radiometer was poorly calibrated following a power outage at the site. |
Measurements: | enamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/26/2014 | 0000 | 10/08/2015 | 1600 | Incorrect |
Subject: | ENA/MWRP/C1 - Reprocess: Calibration drift |
DataStreams: | enamwrpC1.b1 |
Description: | The data are incorrect mostly because of fast drift in the calibration. Brightness temperatures become negative after 9/7/2015. |
Measurements: | enamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/24/2014 | 0200 | 10/08/2015 | 1504 | Suspect |
Subject: | ENA/MWRP/C1 - Occasional outages of surface T/RH readings |
DataStreams: | enamwrpC1.b1 |
Description: | Occasionally the surface temperature and relative humidity readings are not updated and the measurements appear flat for short times. The primary data are most likely not affected because the surface measurements are flatlined for short intervals when big changes are not expected. |
Measurements: | enamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/29/2013 | 0941 | 04/30/2016 | 2359 | Suspect |
Subject: | ENA/C1 - Reprocess: Slightly incorrect location fields |
DataStreams: | enamwr3cC1.b1 |
Description: | Some of the instruments are being processed with the previous "GRW-Azores" lat/lon/alt instead of the new "ENA-Azores" lat/lon/alt. The differences are minor, but will be corrected. The respective GRW and ENA lat, lon, and alt values are as follows: GRW lat = 39.0911 ; ENA lat = 39.0916 ; GRW lon = -28.0297 ; ENA lon = -28.0257 ; GRW alt = 26 ; ENA alt = 30.48 ; |
Measurements: | enamwr3cC1.b1:
|