Data Quality Reports for Session: 185370 User: diaom2 Completed: 05/04/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160427.2AWR/HSRL/M1 - Degraded signalawrhsrlM1.01Incorrect
D160427.4AWR/HSRL/M1 - Hardware issueawrhsrlM1.01Incorrect
D160608.2AWR/HSRL/M1 - Instrument malfunctionawrhsrlM1.01Incorrect


DQRID : D160427.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/06/2016220004/07/20162220Incorrect
Subject:
AWR/HSRL/M1 - Degraded signal
DataStreams:awrhsrlM1.01
Description:
Signal degraded during this period
Measurements:awrhsrlM1.01:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D160427.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/19/2016151001/19/20162120Incorrect
04/01/2016170004/01/20162200Incorrect
Subject:
AWR/HSRL/M1 - Hardware issue
DataStreams:awrhsrlM1.01
Description:
An unknown issue with the hardware caused artifacts in the data.
Measurements:awrhsrlM1.01:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D160608.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/04/2016092306/06/20161100Incorrect
Subject:
AWR/HSRL/M1 - Instrument malfunction
DataStreams:awrhsrlM1.01
Description:
All data collected during this period is bad.  Problems with the output shutter/DAQ/Seed 
voltage were reported by the AMF2 operator. He restarted the computer on June 4th. The 
system lock point was lost at that time. The lock point was manually reset from ~-9V to 2.8V 
on Monday June 6 around 11 UTC.
Measurements:awrhsrlM1.01:
  • Raw data stream - documentation not supported


Back To Table of Contents



END OF DATA