DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160219.4 | ENA/NEPHELOMETER/C1 - Data not saved | enaaosnephwet1mC1.b1 | Missing |
D160613.1 | ENA/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor not switching | enaaosnephwet1mC1.b1 | Suspect |
D160729.7 | ENA/NEPHELOMETER/C1 - Humidigraph flow issues | enaaosnephwet1mC1.b1 | Does not affect quality |
D160909.13 | ENA/NEPHELOMETER/C1 - Inlet flow changed | enaaosnephwet1mC1.b1 | Suspect |
D161117.3 | ENA/NEPHELOMETER/C1 - Instrument offline | enaaosnephwet1mC1.b1 | Missing |
D170428.3 | ENA/HTDMA/NEPHELOMETER/C1 - Dry air supply failing | enaaosnephwet1mC1.b1 | Suspect |
D170428.9 | ENA/HTDMA/NEPHELOMETER/C1 - Dry air supply high RH | enaaosnephwet1mC1.b1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/16/2016 | 1405 | 02/16/2016 | 1718 | Missing |
Subject: | ENA/NEPHELOMETER/C1 - Data not saved |
DataStreams: | enaaosnephwet1mC1.b1 |
Description: | Instrument computer rebooted for unknown reasons and got stuck on booting. No data saved for the period. |
Measurements: | enaaosnephwet1mC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/10/2016 | 1500 | 06/12/2016 | 1100 | Suspect |
Subject: | ENA/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor not switching |
DataStreams: | enaaosnephwet1mC1.b1 |
Description: | Impactor power accidentally unplugged. Impactor not switching and stuck in 10um position. The quality of data variables is not affected. |
Measurements: | enaaosnephwet1mC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/29/2016 | 0929 | 03/02/2017 | 0000 | Does not affect quality |
Subject: | ENA/NEPHELOMETER/C1 - Humidigraph flow issues |
DataStreams: | enaaosnephwet1mC1.b1 |
Description: | Flow issues with the humidigraph. Nephelometers operating at a different RH than specified, but data quality not affected. |
Measurements: | enaaosnephwet1mC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/08/2016 | 0024 | 11/24/2016 | 1730 | Suspect |
Subject: | ENA/NEPHELOMETER/C1 - Inlet flow changed |
DataStreams: | enaaosnephwet1mC1.b1 |
Description: | Due to a failure of another instrument on the same sample line, the sample flow for the nephelometers/humidigraph was different than specified. This potentially affected the humidigraph RH and the impactor cut point. Full effect on data quality is unknown, but is likely limited to the impactor cut point being slightly different than expected. |
Measurements: | enaaosnephwet1mC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/11/2016 | 1200 | 11/14/2016 | 1712 | Missing |
Subject: | ENA/NEPHELOMETER/C1 - Instrument offline |
DataStreams: | enaaosnephwet1mC1.b1 |
Description: | Instrument had a hard drive failure. No data collected during this period. |
Measurements: | enaaosnephwet1mC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/25/2017 | 1100 | 04/26/2017 | 1200 | Suspect |
Subject: | ENA/HTDMA/NEPHELOMETER/C1 - Dry air supply failing |
DataStreams: | enaaosnephwet1mC1.b1 |
Description: | Pentras dry air generator failing and offline for maintenance. Dry air supply to HT-DMA and humidigraph/nephelometers affected. Instruments working well, but with dry RH of around 30% instead <5%. |
Measurements: | enaaosnephwet1mC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/26/2017 | 1100 | 04/28/2017 | 0810 | Suspect |
Subject: | ENA/HTDMA/NEPHELOMETER/C1 - Dry air supply high RH |
DataStreams: | enaaosnephwet1mC1.b1 |
Description: | Humidigraph valve o-ring broken and valve stuck after o-ring replacement. Dry and humid air mixing. Instruments working well, except dry air RH different than nominal. HT-DMA affected, as well, as it is using the same dry air source. |
Measurements: | enaaosnephwet1mC1.b1:
|