Data Quality Reports for Session: 185915 User: pennypackers1 Completed: 05/31/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160621.4ASI/AOS/M1 - Manifold blower not workingasiaoscpcfM1.b1, asiaossmpsM1.a1, asiaoscpcuM1.b1Incorrect
D160708.7ASI/SMPS/M1 - Different neutralizer usedasiaossmpsM1.a1Does not affect quality
D160907.1ASI/CPC/M1 - Collapsed internal tubeasiaoscpcuM1.b1Incorrect
D160907.12ASI/CPC/M1 - Instrument testingasiaoscpcfM1.b1Incorrect
D160907.13ASI/CPC/M1 - Instrument maintenceasiaoscpcfM1.b1Incorrect
D160907.14ASI/SMPS/M1 - Instrument maintenceasiaossmpsM1.a1Incorrect
D161003.1ASI/SMPS/M1 - Small leak in the SMPS CPCasiaossmpsM1.a1Suspect
D161018.2ASI/SMPS/M1/M1 - Instrument computer failureasiaossmpsM1.a1Missing
D161021.3ASI/SMPS/M1 - CollectionasiaossmpsM1.a1Does not affect quality
D161021.4ASI/SMPS/M1 - Different number of particle size binsasiaossmpsM1.a1Does not affect quality
D161024.2ASI/CPC/M1 - CollectionasiaoscpcfM1.b1, asiaoscpcuM1.b1Does not affect quality
D161128.5ASI/CPC/M1 - Saturator wick changed out asiaoscpcuM1.b1Incorrect
D161202.8ASI/CPC/M1 - Counts periodically drifting to zeroasiaoscpcfM1.b1, asiaoscpcuM1.b1Incorrect
D161206.3ASI/CPC/M1 - Saturator wick changed outasiaoscpcfM1.b1Incorrect
D161216.9ASI/CPC/M1 - Debris found in inner nozzleasiaoscpcfM1.b1Incorrect
D170105.10ASI/SMPS/M1 - HEPA Filter TestasiaossmpsM1.a1Incorrect
D170202.10ASI/CPC/SMPS/M1 - ACSM CalibrationasiaossmpsM1.a1, asiaoscpcuM1.b1Missing
D170202.11ASI/SMPS/M1 - Bad Dacron ScreenasiaossmpsM1.a1Incorrect
D170202.12ASI/CPC/SMPS/M1 - ACSM CalibrationasiaossmpsM1.a1, asiaoscpcfM1.b1, asiaoscpcuM1.b1Missing
D170202.13ASI/CPC/M1 - Gillibrator Flow Rate TestasiaoscpcfM1.b1Incorrect
D170202.14ASI/SMPS/M1 - PSL Sizing TestsasiaossmpsM1.a1Incorrect
D170202.15ASI/SMPS/M1 - CCN CalibrationasiaossmpsM1.a1Missing
D170202.16ASI/SMPS/M1 - Software FrozenasiaossmpsM1.a1Missing
D170202.17ASI/SMPS/M1 - HEPA Challenge & Flow RateasiaossmpsM1.a1Incorrect
D170202.18ASI/CPC/M1 - CPCU removed for SMPS/CPCF comparisonasiaoscpcuM1.b1Missing
D170202.19ASI/SMPS/M1 - SMPS/CPCF comparisonasiaossmpsM1.a1Incorrect
D170202.2ASI/SMPS/M1 - Computer Reboot and Leak TestingasiaossmpsM1.a1Incorrect
D170202.20ASI/CPC/M1 - CPCF/SMPS comparisonasiaoscpcfM1.b1Suspect
D170202.21ASI/CPC/M1 - Instrument Thermistor ProblemasiaoscpcuM1.b1Incorrect
D170202.3ASI/SMPS/M1 - HEPA Filter TestsasiaossmpsM1.a1Incorrect
D170202.4ASI/SMPS/M1 - Missing DataasiaossmpsM1.a1Missing
D170202.5ASI/CPC/M1 - Adding quick connect to vacuum lineasiaoscpcfM1.b1Incorrect
D170202.6ASI/CPC/M1 - Adding quick connect to vacuum lineasiaoscpcuM1.b1Incorrect
D170202.7ASI/CPC/UHSAS/M1 - HEPA filter testingasiaoscpcfM1.b1, asiaoscpcuM1.b1Incorrect
D170202.8ASI/CPC/M1 - Cleaning & CalibrationasiaoscpcfM1.b1Missing
D170202.9ASI/CPC/M1 - Cleaning & Calibration CheckasiaoscpcuM1.b1Missing
D170210.5ASI/CPC/M1 - Instrument Butanol fill not initializing, cpc ran dryasiaoscpcfM1.b1Incorrect
D170210.6ASI/CPC/M1 - Instrument Data MissingasiaoscpcfM1.b1Incorrect
D170213.4ASI/CPC/M1 - Debris Blocking CPC Optics BlockasiaoscpcfM1.b1Incorrect
D170214.4ASI/CPC/M1 - Instrument maintenanceasiaoscpcfM1.b1Missing
D170420.4ASI/SMPS/M1 - Software Freeze UpasiaossmpsM1.a1Missing
D170421.2ASI/CPC/M1 - Condenser temperature fluctuatingasiaoscpcuM1.b1Suspect
D170421.4ASI/CPC/M1 - Temperature difference causing high particle number concentrationsasiaoscpcuM1.b1Incorrect
D170421.5ASI/CPC/M1 - Troubleshooting and replacing failed thermistorasiaoscpcfM1.b1Missing
D170421.6ASI/CPC/M1 - InstrumentasiaoscpcfM1.b1, asiaoscpcuM1.b1Incorrect
D170505.10ASI/CPC/M1 - Startup SpikeasiaoscpcuM1.b1Incorrect


DQRID : D160621.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/16/2016004206/16/20161345Incorrect
Subject:
ASI/AOS/M1 - Manifold blower not working
DataStreams:asiaoscpcfM1.b1, asiaossmpsM1.a1, asiaoscpcuM1.b1
Description:
Manifold blower not working, instrument sample flows either zero or different than 
specified. Sampling through the stack not isokinetic.
Measurements:asiaossmpsM1.a1:
  • cpc_inlet_flow
  • base_time
  • DMA_outer_radius
  • lower_size
  • low_voltage
  • diffusion_correction
  • DMA_inner_radius
  • time_offset
  • down_scan_first
  • reference_mean_free_path
  • Total concentration(total_concentration)
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • geometric_mean_std
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • particle_density
  • status_flag
  • Altitude above mean sea level(alt)
  • CPC_sample_flow
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • DMA_characteristic_length
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • sample_pressure
  • cpc_sample_flow
  • median
  • high_voltage
  • d50
  • mode
  • nanoparticle_agglomerate_mobility_analysis
  • CPC_inlet_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • time
  • diameter_midpoint

asiaoscpcuM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • base_time
  • time
  • Altitude above mean sea level(alt)
  • condensation_temp
  • North latitude(lat)
  • cabinet_temp
  • liquid_level
  • qc_concentration
  • East longitude(lon)
  • optics_temp
  • nozzle_pressure
  • aerosol_flow
  • orifice_pressure
  • saturation_temp
  • laser_current
  • ambient_pressure
  • fraction_full
  • time_offset

asiaoscpcfM1.b1:
  • dilution_flow_reading
  • optics_temp
  • base_time
  • nozzle_pressure
  • North latitude(lat)
  • dilution_intercept
  • liquid_level
  • dilution_valve_position
  • time_offset
  • aerosol_flow_calibrated
  • condensation_temp
  • laser_current
  • time
  • dilution_slope
  • cabinet_temp
  • qc_concentration
  • orifice_pressure
  • fraction_full
  • East longitude(lon)
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • saturation_temp
  • ambient_pressure
  • dilution_correction_factor
  • Altitude above mean sea level(alt)


Back To Table of Contents

DQRID : D160708.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/19/2016000007/06/20161535Does not affect quality
Subject:
ASI/SMPS/M1 - Different neutralizer used
DataStreams:asiaossmpsM1.a1
Description:
From the start of sampling up through 07/06/2016, a Po-210 neutralizer was used in lieu of 
 the TSI standard Kr-85 neutralizer.  The Kr-85 neutralizer has now been installed.
Measurements:asiaossmpsM1.a1:
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)


Back To Table of Contents

DQRID : D160907.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/01/2016134509/01/20161422Incorrect
Subject:
ASI/CPC/M1 - Collapsed internal tube
DataStreams:asiaoscpcuM1.b1
Description:
During this time, operator and mentor were troubleshooting a collapsed internal tube on 
the  3776 CPC.
Measurements:asiaoscpcuM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D160907.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/03/2016131409/03/20161327Incorrect
Subject:
ASI/CPC/M1 - Instrument testing
DataStreams:asiaoscpcfM1.b1
Description:
During this time tests were being run with a HEPA filter on the inlet.  Data is not 
representative of ambient aerosol concentrations
Measurements:asiaoscpcfM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D160907.13
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/06/2016100409/06/20161029Incorrect
09/06/2016181509/06/20161819Incorrect
Subject:
ASI/CPC/M1 - Instrument maintence
DataStreams:asiaoscpcfM1.b1
Description:
During these intervals, connections were being removed from the CPC to prepare for it to 
take the place of the SMPS CPC.
Measurements:asiaoscpcfM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D160907.14
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/06/2016103009/06/20161154Incorrect
08/29/2016152708/29/20161549Incorrect
09/06/2016172509/06/20161744Incorrect
08/29/2016160508/31/20161719Incorrect
09/02/2016130509/02/20161334Incorrect
09/06/2016180009/06/20161824Incorrect
09/01/2016130909/01/20161814Incorrect
08/29/2016125508/29/20161344Incorrect
Subject:
ASI/SMPS/M1 - Instrument maintence
DataStreams:asiaossmpsM1.a1
Description:
During these time periods, a HEPA filter was connected to different portions of the system 
for investigation of leaks.
Measurements:asiaossmpsM1.a1:
  • geometric_mean_std
  • mode
  • d50
  • mean
  • geometric_mean
  • Total concentration(total_concentration)
  • median
  • Number size distribution (dN/dlogDp)(number_size_distribution)


Back To Table of Contents

DQRID : D161003.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/29/2016134508/29/20161526Suspect
08/29/2016155008/29/20161604Suspect
09/03/2016123009/03/20161304Suspect
05/19/2016000008/29/20161254Suspect
09/02/2016133509/02/20161419Suspect
Subject:
ASI/SMPS/M1 - Small leak in the SMPS CPC
DataStreams:asiaossmpsM1.a1
Description:
Periodic spikes were seen in data due to a small leak in the SMPS CPC. These spikes are 
identifiable and may be corrected in future data products.
Measurements:asiaossmpsM1.a1:
  • geometric_mean_std
  • mode
  • mean
  • geometric_mean
  • Total concentration(total_concentration)
  • median
  • Number size distribution (dN/dlogDp)(number_size_distribution)


Back To Table of Contents

DQRID : D161018.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/07/2016200010/14/20161054Missing
Subject:
ASI/SMPS/M1/M1 - Instrument computer failure
DataStreams:asiaossmpsM1.a1
Description:
The computer running the SMPS failed.  Some data was lost in the interim during the 
replacement of the computer and some data may be recoverable from the failed computer.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • Altitude above mean sea level(alt)
  • base_time
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • CPC_sample_flow
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • DMA_characteristic_length
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • time_offset
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • CPC_inlet_flow
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • time
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D161021.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/06/2016103009/06/20161809Does not affect quality
Subject:
ASI/SMPS/M1 - Collection
DataStreams:asiaossmpsM1.a1
Description:
The CPC associated with the CPCF measurement was sampling in place of the CPC associated 
with the SMPS measurement.  They are both the same model number.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • CPC_sample_flow
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • DMA_characteristic_length
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • CPC_inlet_flow
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • mean
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D161021.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/29/2016134508/29/20161549Does not affect quality
Subject:
ASI/SMPS/M1 - Different number of particle size bins
DataStreams:asiaossmpsM1.a1
Description:
The data was exported in a different format.  There will be a different number of particle 
size bins during this period.
Measurements:asiaossmpsM1.a1:
  • diameter_midpoint
  • Number size distribution (dN/dlogDp)(number_size_distribution)


Back To Table of Contents

DQRID : D161024.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/06/2016103009/06/20161815Does not affect quality
Subject:
ASI/CPC/M1 - Collection
DataStreams:asiaoscpcfM1.b1, asiaoscpcuM1.b1
Description:
The CPC associated with the SMPS was sampling in place of the CPC associated with this 
CPCF measurement.  The CPC's are the same model number
Measurements:asiaoscpcuM1.b1:
  • cabinet_temp
  • liquid_level
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • optics_temp
  • nozzle_pressure
  • condensation_temp
  • aerosol_flow
  • orifice_pressure
  • saturation_temp
  • laser_current
  • ambient_pressure
  • fraction_full

asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_flow_reading
  • laser_current
  • dilution_slope
  • cabinet_temp
  • optics_temp
  • orifice_pressure
  • fraction_full
  • nozzle_pressure
  • dilution_intercept
  • liquid_level
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • saturation_temp
  • ambient_pressure
  • dilution_correction_factor
  • aerosol_flow_calibrated


Back To Table of Contents

DQRID : D161128.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/21/2016164011/21/20161703Incorrect
Subject:
ASI/CPC/M1 - Saturator wick changed out
DataStreams:asiaoscpcuM1.b1
Description:
The saturator wick on the cpcu was changed out during this period.
Measurements:asiaoscpcuM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D161202.8
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/10/2016212911/13/20161209Incorrect
12/10/2016052412/10/20161418Incorrect
11/25/2016160511/26/20160822Incorrect
11/20/2016192611/21/20161528Incorrect
11/03/2016170911/04/20161206Incorrect
10/26/2016113110/26/20161750Incorrect
12/14/2016061712/14/20161147Incorrect
Subject:
ASI/CPC/M1 - Counts periodically drifting to zero
DataStreams:asiaoscpcfM1.b1, asiaoscpcuM1.b1
Description:
The CPC counts are periodically drifting to zero.  The drift from concentrations 
comparable to the CPCU is somewhat abrupt, but not instantaneous.  Placing a small pressure drop 
on the inlet (e.g. by measuring the flow rate) causes the counts to return to normal.  The 
root cause of the problem has not yet been identified.  More drifts to zero are possible 
in the future.
Measurements:asiaoscpcuM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)

asiaoscpcfM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D161206.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/21/2016174011/21/20161818Incorrect
Subject:
ASI/CPC/M1 - Saturator wick changed out
DataStreams:asiaoscpcfM1.b1
Description:
The saturator wick was changed out at this time.
Measurements:asiaoscpcfM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D161216.9
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/15/2016145512/15/20161706Incorrect
Subject:
ASI/CPC/M1 - Debris found in inner nozzle
DataStreams:asiaoscpcfM1.b1
Description:
During this time, the CPCF was taken offline to clean the inner nozzle.  Debris was found 
in the inner nozzle.  During this time period, the CPC was either sampling cabin air or 
no data is available.
Measurements:asiaoscpcfM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D170105.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/04/2017143501/04/20171449Incorrect
Subject:
ASI/SMPS/M1 - HEPA Filter Test
DataStreams:asiaossmpsM1.a1
Description:
SMPS was sampling through a HEPA filter
Measurements:asiaossmpsM1.a1:
  • geometric_mean_std
  • mode
  • mean
  • geometric_mean
  • Total concentration(total_concentration)
  • median
  • Number size distribution (dN/dlogDp)(number_size_distribution)


Back To Table of Contents

DQRID : D170202.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/23/2017090001/24/20171014Missing
Subject:
ASI/CPC/SMPS/M1 - ACSM Calibration
DataStreams:asiaossmpsM1.a1, asiaoscpcuM1.b1
Description:
The SMPS and CPCU were used to calibrate the ACSM during this time period
Measurements:asiaossmpsM1.a1:
  • cpc_inlet_flow
  • base_time
  • DMA_outer_radius
  • lower_size
  • low_voltage
  • diffusion_correction
  • DMA_inner_radius
  • time_offset
  • down_scan_first
  • reference_mean_free_path
  • Total concentration(total_concentration)
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • geometric_mean_std
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • particle_density
  • status_flag
  • Altitude above mean sea level(alt)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • DMA_characteristic_length
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • sample_pressure
  • cpc_sample_flow
  • median
  • high_voltage
  • d50
  • mode
  • nanoparticle_agglomerate_mobility_analysis
  • reference_gas_temperature
  • reference_gas_viscosity
  • time
  • diameter_midpoint

asiaoscpcuM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • base_time
  • time
  • Altitude above mean sea level(alt)
  • condensation_temp
  • North latitude(lat)
  • cabinet_temp
  • liquid_level
  • qc_concentration
  • East longitude(lon)
  • optics_temp
  • nozzle_pressure
  • aerosol_flow
  • orifice_pressure
  • saturation_temp
  • laser_current
  • ambient_pressure
  • fraction_full
  • time_offset


Back To Table of Contents

DQRID : D170202.11
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/24/2017101501/24/20171159Incorrect
Subject:
ASI/SMPS/M1 - Bad Dacron Screen
DataStreams:asiaossmpsM1.a1
Description:
The size distribution seemed unusual.  Inspection of the DMA revealed a bad dacron screen.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • Altitude above mean sea level(alt)
  • base_time
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • DMA_characteristic_length
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • time_offset
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • time
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/24/2017133501/24/20171644Missing
Subject:
ASI/CPC/SMPS/M1 - ACSM Calibration
DataStreams:asiaossmpsM1.a1, asiaoscpcfM1.b1, asiaoscpcuM1.b1
Description:
The SMPS and CPCU were used to calibrate the ACSM during this time
Measurements:asiaossmpsM1.a1:
  • cpc_inlet_flow
  • base_time
  • DMA_outer_radius
  • lower_size
  • low_voltage
  • diffusion_correction
  • DMA_inner_radius
  • time_offset
  • down_scan_first
  • reference_mean_free_path
  • Total concentration(total_concentration)
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • geometric_mean_std
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • particle_density
  • status_flag
  • Altitude above mean sea level(alt)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • DMA_characteristic_length
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • sample_pressure
  • cpc_sample_flow
  • median
  • high_voltage
  • d50
  • mode
  • nanoparticle_agglomerate_mobility_analysis
  • reference_gas_temperature
  • reference_gas_viscosity
  • time
  • diameter_midpoint

asiaoscpcuM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • base_time
  • time
  • Altitude above mean sea level(alt)
  • condensation_temp
  • North latitude(lat)
  • cabinet_temp
  • liquid_level
  • qc_concentration
  • East longitude(lon)
  • optics_temp
  • nozzle_pressure
  • aerosol_flow
  • orifice_pressure
  • saturation_temp
  • laser_current
  • ambient_pressure
  • fraction_full
  • time_offset

asiaoscpcfM1.b1:
  • dilution_flow_reading
  • optics_temp
  • base_time
  • nozzle_pressure
  • North latitude(lat)
  • dilution_intercept
  • liquid_level
  • dilution_valve_position
  • time_offset
  • aerosol_flow_calibrated
  • condensation_temp
  • laser_current
  • time
  • dilution_slope
  • cabinet_temp
  • qc_concentration
  • orifice_pressure
  • fraction_full
  • East longitude(lon)
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • saturation_temp
  • ambient_pressure
  • dilution_correction_factor
  • Altitude above mean sea level(alt)


Back To Table of Contents

DQRID : D170202.13
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/25/2017102001/25/20171022Incorrect
Subject:
ASI/CPC/M1 - Gillibrator Flow Rate Test
DataStreams:asiaoscpcfM1.b1
Description:
Gillibrator was measuring the flow rate.  CPCF not sampling from the atmosphere.
Measurements:asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_intercept
  • dilution_flow_reading
  • laser_current
  • liquid_level
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • saturation_temp
  • dilution_slope
  • ambient_pressure
  • dilution_correction_factor
  • cabinet_temp
  • optics_temp
  • orifice_pressure
  • fraction_full
  • aerosol_flow_calibrated
  • nozzle_pressure


Back To Table of Contents

DQRID : D170202.14
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/25/2017102501/25/20171109Incorrect
Subject:
ASI/SMPS/M1 - PSL Sizing Tests
DataStreams:asiaossmpsM1.a1
Description:
The SMPS was used to size challenge PSL aerosol during this time.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • DMA_characteristic_length
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • mean
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.15
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/26/2017111501/26/20171554Missing
01/25/2017142001/25/20171539Missing
Subject:
ASI/SMPS/M1 - CCN Calibration
DataStreams:asiaossmpsM1.a1
Description:
The SMPS was used to calibrate the CCN spectrometer during this time.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • base_time
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • DMA_characteristic_length
  • time_offset
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • time
  • mean
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.16
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/26/2017090501/26/20170949Missing
Subject:
ASI/SMPS/M1 - Software Frozen
DataStreams:asiaossmpsM1.a1
Description:
The SMPS AIM software experienced a software freeze-up during this time.  Unsure of the 
root cause.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • Altitude above mean sea level(alt)
  • base_time
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • DMA_inner_radius
  • mean_free_path
  • diffusion_correction
  • DMA_characteristic_length
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • time_offset
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • time
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.17
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/26/2017155501/26/20171604Incorrect
Subject:
ASI/SMPS/M1 - HEPA Challenge & Flow Rate
DataStreams:asiaossmpsM1.a1
Description:
During this time HEPA filter scans and flow rate measurements were taken.  The data is not 
representative of the ambient aerosol size distribution.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • DMA_inner_radius
  • mean_free_path
  • diffusion_correction
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • DMA_characteristic_length
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • mean
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.18
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/27/2017095001/27/20171005Missing
Subject:
ASI/CPC/M1 - CPCU removed for SMPS/CPCF comparison
DataStreams:asiaoscpcuM1.b1
Description:
CPCU removed so SMPS and CPC could take place and be compared side by side with CPCF 
instrument.
Measurements:asiaoscpcuM1.b1:
  • cabinet_temp
  • aerosol_flow
  • liquid_level
  • orifice_pressure
  • saturation_temp
  • qc_concentration
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • East longitude(lon)
  • base_time
  • optics_temp
  • North latitude(lat)
  • laser_current
  • time
  • nozzle_pressure
  • Altitude above mean sea level(alt)
  • ambient_pressure
  • condensation_temp
  • fraction_full
  • time_offset


Back To Table of Contents

DQRID : D170202.19
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/27/2017105501/27/20171104Incorrect
01/27/2017095001/27/20171029Incorrect
Subject:
ASI/SMPS/M1 - SMPS/CPCF comparison
DataStreams:asiaossmpsM1.a1
Description:
The SMPS was taken from normal sampling configuration to be compared side-by-side with the 
CPCF during this period.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • DMA_characteristic_length
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • mean
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/20/2017095001/20/20171154Incorrect
Subject:
ASI/SMPS/M1 - Computer Reboot and Leak Testing
DataStreams:asiaossmpsM1.a1
Description:
Leak testing taking place.  SMPS was not sampling ambient aerosol.
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • Altitude above mean sea level(alt)
  • base_time
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • DMA_characteristic_length
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • time_offset
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • time
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.20
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/27/2017103101/27/20171107Suspect
Subject:
ASI/CPC/M1 - CPCF/SMPS comparison
DataStreams:asiaoscpcfM1.b1
Description:
The CPCF was moved from its normal place to co-sample with the SMPS for a test on the 
inter-comparability of the two instruments during this time. Concentrations do not seem to 
deviate in this sampling configuration
Measurements:asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_intercept
  • dilution_flow_reading
  • laser_current
  • liquid_level
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • saturation_temp
  • dilution_slope
  • ambient_pressure
  • dilution_correction_factor
  • cabinet_temp
  • optics_temp
  • orifice_pressure
  • fraction_full
  • aerosol_flow_calibrated
  • nozzle_pressure


Back To Table of Contents

DQRID : D170202.21
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/28/2017065602/01/20171159Incorrect
Subject:
ASI/CPC/M1 - Instrument Thermistor Problem
DataStreams:asiaoscpcuM1.b1
Description:
The thermistor on the CPCU 3776 malfunctioned during this time. Particle concentrations 
went very high, likely due to homogeneous nucleation of butanol vapor. The thermistor was 
changed and verified to be correct with a thermocouple.
Measurements:asiaoscpcuM1.b1:
  • cabinet_temp
  • laser_current
  • aerosol_flow
  • liquid_level
  • orifice_pressure
  • nozzle_pressure
  • saturation_temp
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • ambient_pressure
  • condensation_temp
  • optics_temp
  • fraction_full


Back To Table of Contents

DQRID : D170202.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/20/2017143001/20/20171719Incorrect
Subject:
ASI/SMPS/M1 - HEPA Filter Tests
DataStreams:asiaossmpsM1.a1
Description:
HEPA filter on the inlet
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • Altitude above mean sea level(alt)
  • base_time
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • DMA_characteristic_length
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • time_offset
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • time
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/21/2017090501/21/20170939Missing
Subject:
ASI/SMPS/M1 - Missing Data
DataStreams:asiaossmpsM1.a1
Description:
Instrument evaluation, no data during this time
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • Altitude above mean sea level(alt)
  • base_time
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • mean_free_path
  • DMA_inner_radius
  • diffusion_correction
  • DMA_characteristic_length
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • time_offset
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • time
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170202.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/21/2017101901/21/20171023Incorrect
Subject:
ASI/CPC/M1 - Adding quick connect to vacuum line
DataStreams:asiaoscpcfM1.b1
Description:
Probable pressure fluctuations during this time
Measurements:asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_intercept
  • laser_current
  • dilution_flow_reading
  • liquid_level
  • time
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • time_offset
  • saturation_temp
  • dilution_slope
  • ambient_pressure
  • dilution_correction_factor
  • optics_temp
  • cabinet_temp
  • qc_concentration
  • base_time
  • orifice_pressure
  • Altitude above mean sea level(alt)
  • fraction_full
  • aerosol_flow_calibrated
  • North latitude(lat)
  • nozzle_pressure
  • East longitude(lon)


Back To Table of Contents

DQRID : D170202.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/21/2017101901/21/20171023Incorrect
Subject:
ASI/CPC/M1 - Adding quick connect to vacuum line
DataStreams:asiaoscpcuM1.b1
Description:
Probable loss in vacuum line pressure during this time
Measurements:asiaoscpcuM1.b1:
  • cabinet_temp
  • aerosol_flow
  • liquid_level
  • orifice_pressure
  • saturation_temp
  • qc_concentration
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • East longitude(lon)
  • base_time
  • optics_temp
  • North latitude(lat)
  • laser_current
  • time
  • nozzle_pressure
  • Altitude above mean sea level(alt)
  • ambient_pressure
  • condensation_temp
  • fraction_full
  • time_offset


Back To Table of Contents

DQRID : D170202.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/21/2017102901/21/20171030Incorrect
01/25/2017101501/25/20171018Incorrect
Subject:
ASI/CPC/UHSAS/M1 - HEPA filter testing
DataStreams:asiaoscpcfM1.b1, asiaoscpcuM1.b1
Description:
HEPA filter on the inlet to all three instruments during this time for zero checks.  All 
instruments went to zero counts
Measurements:asiaoscpcuM1.b1:
  • cabinet_temp
  • liquid_level
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • optics_temp
  • nozzle_pressure
  • condensation_temp
  • aerosol_flow
  • orifice_pressure
  • saturation_temp
  • laser_current
  • ambient_pressure
  • fraction_full

asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_flow_reading
  • laser_current
  • dilution_slope
  • cabinet_temp
  • optics_temp
  • orifice_pressure
  • fraction_full
  • nozzle_pressure
  • dilution_intercept
  • liquid_level
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • saturation_temp
  • ambient_pressure
  • dilution_correction_factor
  • aerosol_flow_calibrated


Back To Table of Contents

DQRID : D170202.8
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/21/2017103501/21/20171048Missing
Subject:
ASI/CPC/M1 - Cleaning & Calibration
DataStreams:asiaoscpcfM1.b1
Description:
CPC was cleaned and evaluated during this time
Measurements:asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_intercept
  • laser_current
  • dilution_flow_reading
  • liquid_level
  • time
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • time_offset
  • saturation_temp
  • dilution_slope
  • ambient_pressure
  • dilution_correction_factor
  • optics_temp
  • cabinet_temp
  • qc_concentration
  • base_time
  • orifice_pressure
  • Altitude above mean sea level(alt)
  • fraction_full
  • aerosol_flow_calibrated
  • North latitude(lat)
  • nozzle_pressure
  • East longitude(lon)


Back To Table of Contents

DQRID : D170202.9
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/21/2017141701/21/20171505Missing
Subject:
ASI/CPC/M1 - Cleaning & Calibration Check
DataStreams:asiaoscpcuM1.b1
Description:
CPCU removed for cleaning and calibration check during this time.
Measurements:asiaoscpcuM1.b1:
  • cabinet_temp
  • aerosol_flow
  • liquid_level
  • orifice_pressure
  • saturation_temp
  • qc_concentration
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • East longitude(lon)
  • base_time
  • optics_temp
  • North latitude(lat)
  • laser_current
  • time
  • nozzle_pressure
  • Altitude above mean sea level(alt)
  • ambient_pressure
  • condensation_temp
  • fraction_full
  • time_offset


Back To Table of Contents

DQRID : D170210.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/08/2017200002/09/20172014Incorrect
Subject:
ASI/CPC/M1 - Instrument Butanol fill not initializing, cpc ran dry
DataStreams:asiaoscpcfM1.b1
Description:
Due to a problem with the ambient pressure sensor, the butanol fill cycle was deactivated. 
 When the butanol did not turn on, eventually the CPC dried out and concentrations 
dropped.
Measurements:asiaoscpcfM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D170210.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/08/2017174402/08/20171751Incorrect
02/08/2017140802/08/20171450Incorrect
Subject:
ASI/CPC/M1 - Instrument Data Missing
DataStreams:asiaoscpcfM1.b1
Description:
Troubleshooting potential problems with the butanol fill solenoid valve and transistors.
Measurements:asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_intercept
  • dilution_flow_reading
  • laser_current
  • liquid_level
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • saturation_temp
  • dilution_slope
  • ambient_pressure
  • dilution_correction_factor
  • cabinet_temp
  • optics_temp
  • orifice_pressure
  • fraction_full
  • aerosol_flow_calibrated
  • nozzle_pressure


Back To Table of Contents

DQRID : D170213.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/13/2017035002/13/20171538Incorrect
Subject:
ASI/CPC/M1 - Debris Blocking CPC Optics Block
DataStreams:asiaoscpcfM1.b1
Description:
During this period there was debris accumulated in the optics block that was affecting 
aerosol counts.
Measurements:asiaoscpcfM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D170214.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/14/2017152902/14/20171604Missing
Subject:
ASI/CPC/M1 - Instrument maintenance
DataStreams:asiaoscpcfM1.b1
Description:
The CPCF was removed for examination of the saturator wick for discoloration.
Measurements:asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_intercept
  • laser_current
  • dilution_flow_reading
  • liquid_level
  • time
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • time_offset
  • saturation_temp
  • dilution_slope
  • ambient_pressure
  • dilution_correction_factor
  • optics_temp
  • cabinet_temp
  • qc_concentration
  • orifice_pressure
  • fraction_full
  • aerosol_flow_calibrated
  • nozzle_pressure


Back To Table of Contents

DQRID : D170420.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/15/2017113003/16/20171129Missing
Subject:
ASI/SMPS/M1 - Software Freeze Up
DataStreams:asiaossmpsM1.a1
Description:
Software froze and stopped collecting data.  It was restarted by site operations
Measurements:asiaossmpsM1.a1:
  • status_flag
  • cpc_inlet_flow
  • Altitude above mean sea level(alt)
  • base_time
  • lower_size
  • DMA_outer_radius
  • low_voltage
  • DMA_inner_radius
  • mean_free_path
  • diffusion_correction
  • DMA_characteristic_length
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • time_offset
  • down_scan_first
  • sheath_flow
  • geometric_mean
  • reference_gas_pressure
  • gas_viscosity
  • reference_mean_free_path
  • sample_pressure
  • Total concentration(total_concentration)
  • cpc_sample_flow
  • median
  • high_voltage
  • mode
  • d50
  • nanoparticle_agglomerate_mobility_analysis
  • Sample temperature(sample_temperature)
  • aerosol_flow
  • reference_gas_temperature
  • reference_gas_viscosity
  • geometric_mean_std
  • time
  • mean
  • North latitude(lat)
  • East longitude(lon)
  • multiple_charge_correction
  • diameter_midpoint
  • particle_density


Back To Table of Contents

DQRID : D170421.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/11/2017100004/13/20172045Suspect
Subject:
ASI/CPC/M1 - Condenser temperature fluctuating
DataStreams:asiaoscpcuM1.b1
Description:
During this time, the condenser temperature was still around the nominal value but was 
fluctuating causing some brief spikes in number concentrations.  The thermistor in the 
condenser temperature feedback control loop eventually failed.
Measurements:asiaoscpcuM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • condensation_temp


Back To Table of Contents

DQRID : D170421.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/13/2017204604/14/20171558Incorrect
04/15/2017165904/17/20171602Incorrect
Subject:
ASI/CPC/M1 - Temperature difference causing high particle number concentrations
DataStreams:asiaoscpcuM1.b1
Description:
The thermistor that maintains the setpoint condenser temperature malfunctioned giving 
erroneous readings.  The condenser was colder than the readout.  The increased temperature 
difference between the condenser and the saturator caused homogeneous nucleation of butanol 
vapor, evidenced by very high particle number concentrations.
Measurements:asiaoscpcuM1.b1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • condensation_temp


Back To Table of Contents

DQRID : D170421.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/17/2017151004/17/20171602Missing
04/14/2017155904/14/20171759Missing
04/14/2017135904/14/20171416Missing
Subject:
ASI/CPC/M1 - Troubleshooting and replacing failed thermistor
DataStreams:asiaoscpcfM1.b1
Description:
The cpcu was taken offline for troubleshooting and replacing the failed thermistor.
Measurements:asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_intercept
  • laser_current
  • dilution_flow_reading
  • liquid_level
  • time
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • time_offset
  • saturation_temp
  • dilution_slope
  • ambient_pressure
  • dilution_correction_factor
  • optics_temp
  • cabinet_temp
  • qc_concentration
  • base_time
  • orifice_pressure
  • Altitude above mean sea level(alt)
  • fraction_full
  • aerosol_flow_calibrated
  • North latitude(lat)
  • nozzle_pressure
  • East longitude(lon)


Back To Table of Contents

DQRID : D170421.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/13/2017000004/17/20170000Incorrect
Subject:
ASI/CPC/M1 - Instrument
DataStreams:asiaoscpcfM1.b1, asiaoscpcuM1.b1
Description:
REQUIRED
Measurements:asiaoscpcuM1.b1:
  • cabinet_temp
  • liquid_level
  • qc_concentration
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • East longitude(lon)
  • base_time
  • optics_temp
  • time
  • nozzle_pressure
  • Altitude above mean sea level(alt)
  • condensation_temp
  • aerosol_flow
  • orifice_pressure
  • saturation_temp
  • North latitude(lat)
  • laser_current
  • ambient_pressure
  • fraction_full
  • time_offset

asiaoscpcfM1.b1:
  • condensation_temp
  • dilution_flow_reading
  • laser_current
  • time
  • dilution_slope
  • cabinet_temp
  • optics_temp
  • qc_concentration
  • base_time
  • orifice_pressure
  • fraction_full
  • North latitude(lat)
  • nozzle_pressure
  • East longitude(lon)
  • dilution_intercept
  • liquid_level
  • dilution_valve_position
  • dilution_flow_setpoint
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • time_offset
  • saturation_temp
  • ambient_pressure
  • dilution_correction_factor
  • Altitude above mean sea level(alt)
  • aerosol_flow_calibrated


Back To Table of Contents

DQRID : D170505.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/03/2017105705/03/20171057Incorrect
Subject:
ASI/CPC/M1 - Startup Spike
DataStreams:asiaoscpcuM1.b1
Description:
The CPC was turned off and when it restarted there was a spike in the data when the 
condenser temperature was lowering to the set point.
Measurements:asiaoscpcuM1.b1:
  • cabinet_temp
  • laser_current
  • aerosol_flow
  • liquid_level
  • orifice_pressure
  • nozzle_pressure
  • saturation_temp
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • ambient_pressure
  • condensation_temp
  • optics_temp
  • fraction_full


Back To Table of Contents



END OF DATA