Data Quality Reports for Session: 185924 User: schenkelsm1 Completed: 05/31/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D170316.1ASI/CEIL/S1 - Reprocess: Incorrect averaging_inverval global attributeasiceilS1.b1Does not affect quality


DQRID : D170316.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/24/2016173011/14/20161559Does not affect quality
Subject:
ASI/CEIL/S1 - Reprocess: Incorrect averaging_inverval global attribute
DataStreams:asiceilS1.b1
Description:
When the ceilometer software was changed from CT-View to CL-View, several firmware 
configuration changes were made to the model CT25K for consistency with the CL31 at other sites, 
including the data acquisition interval which was changed from 15 to 16 seconds. 
However, the global attribute "averaging_interval" in the ceil datastream was not changed.
Measurements:asiceilS1.b1:
  • base_time


Back To Table of Contents



END OF DATA