DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D150302.12 | ACX/CCN/CPC/HTDMA/NEPHELOMETER/OZONE/PSAP/M1 - Ship in port | acxaoso3M1.a1 | Suspect |
D150302.2 | ACX/OZONE/M1 - Data noisy while ship is under way | acxaoso3M1.a1 | Suspect |
D150506.2 | ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit | enaaoso3C1.a1 | Incorrect |
D151116.1 | ENA/OZONE/C1 - Reprocess: Under reporting ozone concentration | enaaoso3C1.a1 | Suspect |
D160427.7 | ENA/C1 - Reprocess: Slightly incorrect location fields | enaaoso3C1.a1 | Suspect |
D160524.3 | SGP/OZONE/S01 - Filter not changed | sgpaoso3S01.a1 | Suspect |
D160610.13 | SGP/S01 - Tower inlet not operational | sgpaoso3S01.a1 | Incorrect |
D160614.11 | AWR/OZONE/M1 - Error changing inlet filter | awraoso3M1.a1 | Incorrect |
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaoso3M1.a1 | Incorrect |
D160622.1 | ASI/OZONE/PTR-MS/TRACEGAS/M1 - Sampling cabin air | asiaoso3M1.a1 | Incorrect |
D160705.1 | ENA/OZONE/C1 - Air inlet plugged | enaaoso3C1.a1 | Does not affect quality |
D160818.2 | ENA/OZONE/C1 - Failing pump | enaaoso3C1.a1 | Does not affect quality |
D160824.5 | ENA/OZONE/C1 - Reprocessed and Corrected: enaosozoneC1.a1 to enaaoso3C1.a1 | enaaoso3C1.a1 | Does not affect quality |
D160824.8 | SGP/OZONE/S01 - Reprocessed and Corrected: sgpaosozoneS01.a1 to sgpaoso3S01.a1 | sgpaoso3S01.a1 | Does not affect quality |
D160901.1 | AWR/OZONE/M1 - Reprocessed and Corrected: awraosozoneM1.a1 to awraoso3M1.a1 | awraoso3M1.a1 | Does not affect quality |
D160909.17 | ENA/OZONE/C1 - Water condensing in the line | enaaoso3C1.a1 | Incorrect |
D161025.6 | ENA/AOS/C1 - Instrument mentors working at the site | enaaoso3C1.a1 | Suspect |
D161026.2 | ENA/AOSMET/OZONE/TRACEGAS/C1 - missing data | enaaoso3C1.a1 | Missing |
D161109.2 | ASI/OZONE/M1 - Not sampling outside air | asiaoso3M1.a1 | Incorrect |
D161117.4 | SGP/TRACEGAS/S01 - Instrument | sgpaoso3S01.a1 | Missing |
D161215.6 | ENA/OZONE/C1 - missing data and extra calibrations | enaaoso3C1.a1 | Does not affect quality |
D170104.2 | ACX/OZONE/M1 - Reprocessed and Corrected: acxaosozoneM1.a1 to acxaoso3M1.a1 | acxaoso3M1.a1 | Does not affect quality |
D170104.3 | ASI/OZONE/M1 - Reprocessed and Corrected: asiaosozoneM1.a1 to asiaoso3M1.a1 | asiaoso3M1.a1 | Does not affect quality |
D170208.3 | ASI/OZONE/TRACEGAS/M1 - Trace gas inlet blower failure | asiaoso3M1.a1 | Incorrect |
D170208.4 | ASI/OZONE/TRACEGAS/M1 - Instrument computer failure | asiaoso3M1.a1 | Missing |
D170303.12 | ENA/AOS/C1 - Dry air supply intermittently not working | enaaoso3C1.a1 | Does not affect quality |
D170426.2 | ASI/OZONE/TRACEGAS/M1 - Inlet blower replacement | asiaoso3M1.a1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/29/2015 | 1748 | 01/30/2015 | 1805 | Suspect |
Subject: | ACX/CCN/CPC/HTDMA/NEPHELOMETER/OZONE/PSAP/M1 - Ship in port |
DataStreams: | acxaoso3M1.a1 |
Description: | Ship was in port and data were likely impacted by anthropogenic sources. |
Measurements: | acxaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/09/2015 | 0000 | 02/16/2015 | 0000 | Suspect |
Subject: | ACX/OZONE/M1 - Data noisy while ship is under way |
DataStreams: | acxaoso3M1.a1 |
Description: | Data is noisy at high frequency while ship is under way. Regular daily calibrations show the accuracy is correct, only the noise is severe (+/- 5-10 ppbv). |
Measurements: | acxaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/22/2015 | 1700 | 04/27/2015 | 1700 | Incorrect |
Subject: | ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit |
DataStreams: | enaaoso3C1.a1 |
Description: | There was visit of 7 AOS instrument mentors to the ENA AOS during this period of instrument repair, calibration and modification. Instruments were occasionally off, or with no sample flow, or receiving calibration samples, etc. There was no productive sampling of ambient air during this period other than for testing. No data should be used for this period as ambient samples. Note: Two datastream are missing. The CAPS Extinction Monitor and the Greenhouse gas instrument. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/17/2015 | 0000 | 01/25/2016 | 1436 | Suspect |
Subject: | ENA/OZONE/C1 - Reprocess: Under reporting ozone concentration |
DataStreams: | enaaoso3C1.a1 |
Description: | Switching valve developed an internal leak that gradually worsened. Reported instrument values are under-representations of the ambient ozone concentration during this period. The under reporting worsens from 85% to 77% over the course of this DQR. No corrections have been applied at this point. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/29/2013 | 0941 | 04/30/2016 | 2359 | Suspect |
Subject: | ENA/C1 - Reprocess: Slightly incorrect location fields |
DataStreams: | enaaoso3C1.a1 |
Description: | Some of the instruments are being processed with the previous "GRW-Azores" lat/lon/alt instead of the new "ENA-Azores" lat/lon/alt. The differences are minor, but will be corrected. The respective GRW and ENA lat, lon, and alt values are as follows: GRW lat = 39.0911 ; ENA lat = 39.0916 ; GRW lon = -28.0297 ; ENA lon = -28.0257 ; GRW alt = 26 ; ENA alt = 30.48 ; |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/25/2016 | 0000 | 04/19/2016 | 1744 | Suspect |
Subject: | SGP/OZONE/S01 - Filter not changed |
DataStreams: | sgpaoso3S01.a1 |
Description: | Filter not changed. Pressure low and response to calibrations increased. The time before the filter change is considered suspect. |
Measurements: | sgpaoso3S01.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/05/2016 | 1423 | 05/05/2016 | 1740 | Incorrect |
Subject: | SGP/S01 - Tower inlet not operational |
DataStreams: | sgpaoso3S01.a1 |
Description: | Inlet not operational in order to change met sensor. |
Measurements: | sgpaoso3S01.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/12/2016 | 0019 | 01/26/2016 | 0601 | Incorrect |
Subject: | AWR/OZONE/M1 - Error changing inlet filter |
DataStreams: | awraoso3M1.a1 |
Description: | Operator mistakenly used blue filter separator instead of white filter. No flow through instrument. |
Measurements: | awraoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2016 | 0042 | 06/16/2016 | 1345 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
DataStreams: | asiaoso3M1.a1 |
Description: | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements: | asiaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/25/2016 | 0837 | 05/25/2016 | 1251 | Incorrect |
Subject: | ASI/OZONE/PTR-MS/TRACEGAS/M1 - Sampling cabin air |
DataStreams: | asiaoso3M1.a1 |
Description: | Inlet system turned off. Sampling cabin air. No effect on NO/NOx/NOy. |
Measurements: | asiaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/10/2016 | 1215 | 07/05/2016 | 1910 | Does not affect quality |
Subject: | ENA/OZONE/C1 - Air inlet plugged |
DataStreams: | enaaoso3C1.a1 |
Description: | The zero air inlet was plugged after a leak test. Mentor failed to identify the problem until 2016-07-05. Data should be valid. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/22/2016 | 1400 | 09/20/2016 | 1100 | Does not affect quality |
Subject: | ENA/OZONE/C1 - Failing pump |
DataStreams: | enaaoso3C1.a1 |
Description: | The pump was failing, and the operator attempted a repair. Eventually, the flow bypassed the pump internally and was replaced with an external HTDMA vacuum pump. Data appear valid since the flow only affects the internal calibration source and not the ambient measurement. Internal pump was repaired on 2016-09-20 11:00. The flow returned to the nominal value decreased slightly). The response to calibration increased proportionally (as expected given the lower flow rate and longer residence time in the O3 generator cell). The instrument response to ambient O3 should not be a function of the flow rate. HOWEVER, a slight discontinuity was noted in the ambient O3 values prior to this DQR and following this DQR end date. The amount of uncertainty was judged as <5 ppbv. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2013 | 0000 | 11/02/2016 | 1700 | Does not affect quality |
Subject: | ENA/OZONE/C1 - Reprocessed and Corrected: enaosozoneC1.a1 to enaaoso3C1.a1 |
DataStreams: | enaaoso3C1.a1 |
Description: | DATA REPROCESSED and CORRECTED: The enaaosozone.a1 data was renamed to enaaoso3.a1. In addition to the rename, we added data from raw data columns that were originally not included. These additional columns will be used in subsequent downstream processing. The data were archived February 2017. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-02 REPROCESS IDENTIFICATION NUMBER: RID 1493 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The enaaosozoneC1.a1 data is being renamed to aoso3.a1. In addition to the rename, we are adding data from raw data columns that were missed. These additional columns will be used in subsequent downstream processing. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/02/2016 | 0000 | 06/10/2016 | 2359 | Does not affect quality |
Subject: | SGP/OZONE/S01 - Reprocessed and Corrected: sgpaosozoneS01.a1 to sgpaoso3S01.a1 |
DataStreams: | sgpaoso3S01.a1 |
Description: | DATA REPROCESSED and CORRECTED: The aosozone.a1 data was renamed to aoso3.a1. In addition to the rename, we added data from raw data columns that were originally not included. These additional columns will be used in subsequent downstream processing. The data were archived February 2017. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-02 REPROCESS IDENTIFICATION NUMBER: RID 1496 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The sgpaosozoneA01.a1 data is being renamed to aoso3.a1. In addition to the rename, we are adding data from raw data columns that were missed. These additional columns will be used in subsequent downstream processing. |
Measurements: | sgpaoso3S01.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/18/2015 | 0000 | 11/02/2016 | 1700 | Does not affect quality |
Subject: | AWR/OZONE/M1 - Reprocessed and Corrected: awraosozoneM1.a1 to awraoso3M1.a1 |
DataStreams: | awraoso3M1.a1 |
Description: | DATA REPROCESSED and CORRECTED: The awraosozone.a1 data was renamed to awraoso3.a1. In addition to the rename, we added data from raw data columns that were originally not included. These additional columns will be used in subsequent downstream processing. The data were archived February 2017. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-02 REPROCESS IDENTIFICATION NUMBER: RID 1498 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The awraosozoneM1.a1 data is being renamed to aoso3.a1. In addition to the rename, we are adding data from raw data columns that were missed. These additional columns will be used in subsequent downstream processing. |
Measurements: | awraoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/17/2016 | 1035 | 08/18/2016 | 0538 | Incorrect |
08/15/2016 | 0202 | 08/15/2016 | 0939 | Incorrect |
08/20/2016 | 1348 | 08/21/2016 | 1726 | Incorrect |
08/24/2016 | 0840 | 08/24/2016 | 1105 | Incorrect |
08/01/2016 | 0000 | 08/04/2016 | 1832 | Incorrect |
Subject: | ENA/OZONE/C1 - Water condensing in the line |
DataStreams: | enaaoso3C1.a1 |
Description: | Instrument suffered major oscillations in the signal. Working with Operator it was discovered that the Split System Air Conditioner was blowing directly on the Ozone detector. Water was condensing in the line. The air was redirected and the temperature setpoint raised. This reduced but didn't entirely eliminate the oscillations. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/26/2016 | 0830 | 10/01/2016 | 0100 | Suspect |
Subject: | ENA/AOS/C1 - Instrument mentors working at the site |
DataStreams: | enaaoso3C1.a1 |
Description: | Instrument mentors working at the site. Sample lines and electrical connections being reconfigured. Instruments that had no work done on them may still be affected. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/25/2016 | 0050 | 10/25/2016 | 0518 | Missing |
Subject: | ENA/AOSMET/OZONE/TRACEGAS/C1 - missing data |
DataStreams: | enaaoso3C1.a1 |
Description: | The Unit 1 computer was found stuck in boot screen and had to be powered off to reboot. Data is missing for this period. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/25/2016 | 2046 | 10/26/2016 | 1135 | Incorrect |
Subject: | ASI/OZONE/M1 - Not sampling outside air |
DataStreams: | asiaoso3M1.a1 |
Description: | Inlet turned off (or the blower turned off). So sampling was not on outside air. |
Measurements: | asiaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/17/2016 | 1021 | 04/18/2016 | 1415 | Missing |
03/27/2016 | 0834 | 03/28/2016 | 2000 | Missing |
Subject: | SGP/TRACEGAS/S01 - Instrument |
DataStreams: | sgpaoso3S01.a1 |
Description: | Improper start up following a power failure. Computers came back up but instruments did not restart (by design for Oliktok). Daily checklists not compiled this period (weekend). |
Measurements: | sgpaoso3S01.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/15/2016 | 1200 | 12/15/2016 | 1832 | Does not affect quality |
Subject: | ENA/OZONE/C1 - missing data and extra calibrations |
DataStreams: | enaaoso3C1.a1 |
Description: | The 49i ozone instrument in ENA has difficulty with maintaining the data connection and the data stream while receiving the commands to do zeros and calibrations and has had gaps in the data that interfere with data processing. During the time range a new version of the software that should be better at recovering from the errors was being tested. There is a 2 hour gap in the data and a section of time with zeros and cals every 20 minutes. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/09/2015 | 1900 | 02/12/2015 | 2359 | Does not affect quality |
Subject: | ACX/OZONE/M1 - Reprocessed and Corrected: acxaosozoneM1.a1 to acxaoso3M1.a1 |
DataStreams: | acxaoso3M1.a1 |
Description: | DATA REPROCESSED and CORRECTED: The aosozone.a1 data was renamed to aoso3.a1. In addition to the rename, we added data from raw data columns that were originally not included. These additional columns will be used in subsequent downstream processing. The data were archived February 2017. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-02 REPROCESS IDENTIFICATION NUMBER: RID 1525 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The aosozone.a1 data is being renamed to aoso3.a1. In addition to the rename, we are adding data from raw data columns that were missed. These additional columns will be used in subsequent downstream processing. |
Measurements: | acxaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/23/2016 | 1156 | 11/02/2016 | 1700 | Does not affect quality |
Subject: | ASI/OZONE/M1 - Reprocessed and Corrected: asiaosozoneM1.a1 to asiaoso3M1.a1 |
DataStreams: | asiaoso3M1.a1 |
Description: | DATA REPROCESSED and CORRECTED: The aosozone.a1 data was renamed to aoso3.a1. In addition to the rename, we added data from raw data columns that were originally not included. These additional columns will be used in subsequent downstream processing. The data were archived February 2017. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-02 REPROCESS IDENTIFICATION NUMBER: RID 1526 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The aosozone.a1 data is being renamed to aoso3.a1. In addition to the rename, we are adding data from raw data columns that were missed. These additional columns will be used in subsequent downstream processing. |
Measurements: | asiaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/21/2016 | 2046 | 11/22/2016 | 1300 | Incorrect |
Subject: | ASI/OZONE/TRACEGAS/M1 - Trace gas inlet blower failure |
DataStreams: | asiaoso3M1.a1 |
Description: | MAOS C Trace gas inlet had the blower die. Though there are numbers, they should be viewed as incorrect. It also affects the PTR-MS. |
Measurements: | asiaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/14/2016 | 0157 | 12/14/2016 | 0951 | Missing |
Subject: | ASI/OZONE/TRACEGAS/M1 - Instrument computer failure |
DataStreams: | asiaoso3M1.a1 |
Description: | Unit three computer failed. No data recorded until computer rebuilt remotely. |
Measurements: | asiaoso3M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/20/2016 | 0000 | 03/03/2017 | 0000 | Does not affect quality |
Subject: | ENA/AOS/C1 - Dry air supply intermittently not working |
DataStreams: | enaaoso3C1.a1 |
Description: | Pentras dry air supply intermittently not working. Dry air missing for the humidigraph and the HT-DMA. Dry air supply for those instruments partially replaced by regular undried compressed air, lowering the delivery pressure. No direct effect on other instruments seen. |
Measurements: | enaaoso3C1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/24/2017 | 0403 | 02/24/2017 | 1528 | Incorrect |
Subject: | ASI/OZONE/TRACEGAS/M1 - Inlet blower replacement |
DataStreams: | asiaoso3M1.a1 |
Description: | Bypass blower off (and replaced) |
Measurements: | asiaoso3M1.a1:
|