DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160323.3 | OLI/MASC/M1 - Missing instrument data possible | olimascM1.a1 | Suspect |
D170110.1 | OLI/M1 - Severe Storm | olimascM1.a1 | Does not affect quality |
D170217.2 | OLI/MASC/M1 - Collection | olimascM1.a1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/25/2015 | 0000 | 08/25/2016 | 2359 | Suspect |
Subject: | OLI/MASC/M1 - Missing instrument data possible |
DataStreams: | olimascM1.a1 |
Description: | The MASC instrument has been installed on a pole on the corner of a connex container roof. The location is preliminary for this season, it causes aerodynamic problems. During windy situations (almost all the time during snowfall up in Oliktok) there is probably a kind of swirl forming and the snowflakes are concentrated around the margins of the inlet ring. Triggering only occurs when snowflakes pass close to the center of the ring. The location atop the container is obstructed and results in a loss of data. Double fence wind shielding is proposed for the following seasons. A move of the MASC to an unobstructed location away from the containers and installation of wind shielding is planned once the weather conditions are more favorable the latest in fall 2016. |
Measurements: | olimascM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/05/2017 | 1000 | 01/08/2017 | 1200 | Does not affect quality |
Subject: | OLI/M1 - Severe Storm |
DataStreams: | olimascM1.a1 |
Description: | The site lost power during a severe winter storm. Operators on site were not able to get out and restart the generators to bring things back up until 1/7/2017. Most of the instruments were brought up at this time, but some problems remained until 1/8/2017 |
Measurements: | olimascM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/11/2017 | 2237 | 01/24/2017 | 2200 | Missing |
Subject: | OLI/MASC/M1 - Collection |
DataStreams: | olimascM1.a1 |
Description: | A MASC snowflake camera software problem occurred causing failure to save MASC snowflake imagery to the computer disk. Consequently image collection is missing during this period. The software was restarted and tested, and image data collection resumed January 24, 2017. |
Measurements: | olimascM1.a1:
|