DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D141017.4 | MAO/CPC/S1 - CPCU under-counting | maoaoscpcuS1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/04/2014 | 1700 | 03/25/2014 | 1800 | Incorrect |
09/12/2014 | 1223 | 10/01/2014 | 1600 | Incorrect |
10/30/2014 | 1630 | 11/04/2014 | 1600 | Incorrect |
08/25/2014 | 0200 | 08/25/2014 | 2330 | Incorrect |
08/14/2014 | 2300 | 08/19/2014 | 2200 | Incorrect |
08/20/2014 | 1000 | 08/22/2014 | 1915 | Incorrect |
05/18/2014 | 1800 | 07/22/2014 | 1800 | Incorrect |
01/24/2014 | 1130 | 02/03/2014 | 0100 | Incorrect |
Subject: | MAO/CPC/S1 - CPCU under-counting |
DataStreams: | maoaoscpcuS1.b1 |
Description: | Under normal operating conditions, the CPCU measured number concentration should be equal to or greater than the CPCF measured number concentration. The fact that the CPCU is under-counting with respect to the CPCF during this time period strongly suggests that there is a clog in the sampling capillary, resulting in a lower than expected sample flow rate and therefore a lower than expected measured number concentration. |
Measurements: | maoaoscpcuS1.b1:
|