Data Quality Reports for Session: 186024 User: cirinog1 Completed: 06/06/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D140530.2MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inletmaoaossmpsS1.a1Suspect
D140815.1MAO/AOS/S1/M1 - Instrument maintenancemaoaossmpsS1.a1Incorrect
D141203.3MAO/ALL/S1 - Inlet down/off for cleaning and maintenance.maoaossmpsS1.a1Incorrect
D150311.17MAO/SMPS/S1 - Instrument maintenancemaoaossmpsS1.a1Incorrect
D150429.6MAO/AOS/S1 - Flow rate changemaoaossmpsS1.a1Does not affect quality
D150619.7MAO/ACSM/AETH/CCN/CPC/HTDMA/NEPHELOMETER/PASS/PSAP/SMPS/SP2/UHSAS/S1 - Manifold blower
stopped
maoaossmpsS1.a1Incorrect
D150722.4MAO/SMPS/S1 - Missing DatamaoaossmpsS1.a1Missing
D150917.6MAO/ACSM/AETH/CPC/NEPHELOMETER/PSAP/SMPS/UHSAS/S1 Calibration of the NephelometermaoaossmpsS1.a1Incorrect
D150921.2MAO/AOS/M1/S1 - Forest FiremaoaossmpsS1.a1Does not affect quality
D150928.1MAO/SMPS/S1 - Data not availablemaoaossmpsS1.a1Missing
D150928.3MAO/SMPS/S1 - Calibration of SMPSmaoaossmpsS1.a1Incorrect
D150928.5MAO/CPC/SMPS/S1 - Sampling cabin airmaoaossmpsS1.a1Incorrect
D150928.6MAO/SMPS/S1 - Insects clogged nafion dryer and SMPS impactormaoaossmpsS1.a1Suspect
D151203.4MAO/SMPS/S1 - Reprocess: SMPS datamaoaossmpsS1.a1Does not affect quality


DQRID : D140530.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/07/2014143005/13/20142043Suspect
Subject:
MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inlet
DataStreams:maoaossmpsS1.a1
Description:
The clogging of the impactor leads to a decrease in the aerosol sample flow rate of the 
SMPS, which is nominally 1 lpm.  This decrease in the SMPS sample flow rate results in an 
underestimation of the ambient size distribution and integrated number concentration.
Measurements:maoaossmpsS1.a1:
  • aerosol_flow
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • CPC_sample_flow


Back To Table of Contents

DQRID : D140815.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/09/2014160008/15/20141500Incorrect
Subject:
MAO/AOS/S1/M1 - Instrument maintenance
DataStreams:maoaossmpsS1.a1
Description:
Instrument maintenance and cleaning inlets affecting data.
Measurements:maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • sample_pressure
  • geometric_mean_std
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D141203.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/10/2014132508/10/20142115Incorrect
Subject:
MAO/ALL/S1 - Inlet down/off for cleaning and maintenance.
DataStreams:maoaossmpsS1.a1
Description:
Inlet stack of MAOS A lowered. All instruments on internal cabin air. No atmospheric 
validity on HTDMA, PSAP, f(RH) dry neph, wet neph, aethalometer, PASS3, SP2, ACSM, CCN-200, 
Impactor, CPC, UCPC, SMPS.
Measurements:maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • sample_pressure
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D150311.17
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/06/2015155803/06/20151612Incorrect
03/05/2015145803/05/20151522Incorrect
03/04/2015134503/04/20151348Incorrect
03/05/2015145803/06/20151848Incorrect
03/05/2015135103/05/20151414Incorrect
03/04/2015130203/04/20151317Incorrect
03/05/2015143503/05/20151458Incorrect
Subject:
MAO/SMPS/S1 - Instrument maintenance
DataStreams:maoaossmpsS1.a1
Description:
A number of instrument maintenance tasks were performed.  The flow rate of the system from 
the impactor was tested.  Inlet flow rates and nozzle pressures were examined on the CPC 
3772.  The accuracy of the classifier voltage and sheath flow rates were evaluated.  A 
HEPA filter zero test was performed.  During an approximately 24 hour period, the voltage 
cable was left unplugged from the classifier.
Measurements:maoaossmpsS1.a1:
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • high_voltage
  • mean
  • d50
  • Total concentration(total_concentration)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • mode


Back To Table of Contents

DQRID : D150429.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/17/2014130004/20/20151845Does not affect quality
Subject:
MAO/AOS/S1 - Flow rate change
DataStreams:maoaossmpsS1.a1
Description:
At begining of period, the bypass blower on MAOS A failed.  This was replaced by the 
instrument operator, but operator was unable to reprogram the PID flow controller.  The blower 
ran approximately 50% faster than designed flow rate.  It is unlikely this significantly 
affected the flow dynamics or transmission efficiencies in the MAOS A sampling stack.  
The data is colored transparent, but the DQR is filed to alert the user.  No specific 
degradation of signals were noted at this time.
Measurements:maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • sample_pressure
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D150619.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/16/2015150606/17/20151321Incorrect
Subject:
MAO/ACSM/AETH/CCN/CPC/HTDMA/NEPHELOMETER/PASS/PSAP/SMPS/SP2/UHSAS/S1 - Manifold blower 
stopped
DataStreams:maoaossmpsS1.a1
Description:
Manifold blower pulls air through the inlet to the instrument.  Without the blower, the 
air source is indeterminant.
Measurements:maoaossmpsS1.a1:
  • East longitude(lon)
  • low_voltage
  • time
  • mean
  • sheath_flow
  • Altitude above mean sea level(alt)
  • d50
  • aerosol_flow
  • mean_free_path
  • CPC_inlet_flow
  • status_flag
  • mode
  • time_offset
  • CPC_sample_flow
  • gas_viscosity
  • median
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • base_time
  • North latitude(lat)
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • sample_pressure
  • geometric_mean_std


Back To Table of Contents

DQRID : D150722.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/03/2015140807/03/20151552Missing
Subject:
MAO/SMPS/S1 - Missing Data
DataStreams:maoaossmpsS1.a1
Description:
The Electro Static Classifier and Long DMA were used to size aerosol for a calibration of 
the SP2.
Measurements:maoaossmpsS1.a1:
  • East longitude(lon)
  • low_voltage
  • time
  • mean
  • sheath_flow
  • Altitude above mean sea level(alt)
  • d50
  • aerosol_flow
  • mean_free_path
  • CPC_inlet_flow
  • status_flag
  • mode
  • time_offset
  • CPC_sample_flow
  • gas_viscosity
  • median
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • base_time
  • North latitude(lat)
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • sample_pressure
  • geometric_mean_std


Back To Table of Contents

DQRID : D150917.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/17/2015193409/17/20152035Incorrect
Subject:
MAO/ACSM/AETH/CPC/NEPHELOMETER/PSAP/SMPS/UHSAS/S1  Calibration of the Nephelometer
DataStreams:maoaossmpsS1.a1
Description:
Inlet line for these instruments was used during Nephelometer calibration compromising the 
data.
Measurements:maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • sample_pressure
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D150921.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/04/2015000010/08/20152359Does not affect quality
09/13/2015165309/13/20151810Does not affect quality
09/03/2015180009/03/20152359Does not affect quality
Subject:
MAO/AOS/M1/S1 - Forest Fire
DataStreams:maoaossmpsS1.a1
Description:
Smoke from forest burning can be seen on the TSI images rising and then subsequently 
moving over the site during this period.  Many of the AOS instruments observed a sharp spike 
in data values.

October 4-8 multiple fires in the area resulted in hazy conditions during this entire 
period.

Most notable in the following instruments and variables

ACSM - Total Organics
AETH - Black Carbon Concentrations at all wavelengths
CCN100 - Spike in number concentrations occurred a few hours after the smoke plume
CLAP3W - Absorption Coefficient
CPC - Similar to the CCN100, spike in particle concentration occurred a few hours later
CPCF - Particle Concentration
CPCU - Particle Concentration
NEPHELOMETERS - Scattering/Backscattering Coeff
PSAP3W - Absorption Coefficient
SMPS - Concentration and Size Distribution
UHSAS - Size Distribution
Measurements:maoaossmpsS1.a1:
  • median
  • geometric_mean
  • mean_free_path
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • geometric_mean_std
  • mode
  • mean


Back To Table of Contents

DQRID : D150928.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/16/2015121009/16/20151316Missing
09/15/2015175009/15/20152057Missing
Subject:
MAO/SMPS/S1 - Data not available
DataStreams:maoaossmpsS1.a1
Description:
The SMPS was being used as a calibration device for the calibration of the ACSM during 
these periods and no samples were taken.
Measurements:maoaossmpsS1.a1:
  • East longitude(lon)
  • low_voltage
  • time
  • mean
  • sheath_flow
  • Altitude above mean sea level(alt)
  • d50
  • aerosol_flow
  • mean_free_path
  • CPC_inlet_flow
  • status_flag
  • mode
  • time_offset
  • CPC_sample_flow
  • gas_viscosity
  • median
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • base_time
  • North latitude(lat)
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • sample_pressure
  • geometric_mean_std


Back To Table of Contents

DQRID : D150928.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/18/2015185509/18/20151922Incorrect
09/17/2015183609/17/20151852Incorrect
09/17/2015123309/17/20151331Incorrect
09/16/2015140709/16/20152048Incorrect
09/18/2015192909/18/20151949Incorrect
Subject:
MAO/SMPS/S1 - Calibration of SMPS
DataStreams:maoaossmpsS1.a1
Description:
During the times listed, the SMPS was undergoing procedures for calibration of both the 
electrostatic classifier and the condensation particle counter.  Any data that may have 
been collected during this time is not representative of the conditions present in the 
atmosphere at the site.
Measurements:maoaossmpsS1.a1:
  • East longitude(lon)
  • low_voltage
  • time
  • mean
  • sheath_flow
  • Altitude above mean sea level(alt)
  • d50
  • aerosol_flow
  • mean_free_path
  • CPC_inlet_flow
  • status_flag
  • mode
  • time_offset
  • CPC_sample_flow
  • gas_viscosity
  • median
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • base_time
  • North latitude(lat)
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • sample_pressure
  • geometric_mean_std


Back To Table of Contents

DQRID : D150928.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/18/2015151809/18/20151655Incorrect
Subject:
MAO/CPC/SMPS/S1 - Sampling cabin air
DataStreams:maoaossmpsS1.a1
Description:
The SMPS, CPCF, and CPCU were all sampling air from the cabin and not through the aerosol 
inlet during this time.
Measurements:maoaossmpsS1.a1:
  • Number size distribution (dN/dlogDp)(number_size_distribution)


Back To Table of Contents

DQRID : D150928.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/15/2015205709/19/20151408Suspect
Subject:
MAO/SMPS/S1 - Insects clogged nafion dryer and SMPS impactor
DataStreams:maoaossmpsS1.a1
Description:
During this time period the SMPS total concentration was low relative to the cpcf and cpcu 
concentrations.  Upon troubleshooting and examination of components it was discovered 
that a number of small insects had clogged the nafion dryer and SMPS impactor.  The counts 
may have been lowered due to additional losses from bugs.
Measurements:maoaossmpsS1.a1:
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • sample_pressure
  • CPC_sample_flow


Back To Table of Contents

DQRID : D151203.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/29/2014000012/28/20150000Does not affect quality
Subject:
MAO/SMPS/S1 - Reprocess: SMPS data
DataStreams:maoaossmpsS1.a1
Description:
No affect on data quality
Measurements:maoaossmpsS1.a1:
  • status_flag


Back To Table of Contents



END OF DATA