DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140530.2 | MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inlet | maoaossmpsS1.a1 | Suspect |
D140815.1 | MAO/AOS/S1/M1 - Instrument maintenance | maoaossmpsS1.a1 | Incorrect |
D141203.3 | MAO/ALL/S1 - Inlet down/off for cleaning and maintenance. | maoaossmpsS1.a1 | Incorrect |
D150429.6 | MAO/AOS/S1 - Flow rate change | maoaossmpsS1.a1 | Does not affect quality |
D151203.4 | MAO/SMPS/S1 - Reprocess: SMPS data | maoaossmpsS1.a1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/07/2014 | 1430 | 05/13/2014 | 2043 | Suspect |
Subject: | MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inlet |
DataStreams: | maoaossmpsS1.a1 |
Description: | The clogging of the impactor leads to a decrease in the aerosol sample flow rate of the SMPS, which is nominally 1 lpm. This decrease in the SMPS sample flow rate results in an underestimation of the ambient size distribution and integrated number concentration. |
Measurements: | maoaossmpsS1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/09/2014 | 1600 | 08/15/2014 | 1500 | Incorrect |
Subject: | MAO/AOS/S1/M1 - Instrument maintenance |
DataStreams: | maoaossmpsS1.a1 |
Description: | Instrument maintenance and cleaning inlets affecting data. |
Measurements: | maoaossmpsS1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/10/2014 | 1325 | 08/10/2014 | 2115 | Incorrect |
Subject: | MAO/ALL/S1 - Inlet down/off for cleaning and maintenance. |
DataStreams: | maoaossmpsS1.a1 |
Description: | Inlet stack of MAOS A lowered. All instruments on internal cabin air. No atmospheric validity on HTDMA, PSAP, f(RH) dry neph, wet neph, aethalometer, PASS3, SP2, ACSM, CCN-200, Impactor, CPC, UCPC, SMPS. |
Measurements: | maoaossmpsS1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/17/2014 | 1300 | 04/20/2015 | 1845 | Does not affect quality |
Subject: | MAO/AOS/S1 - Flow rate change |
DataStreams: | maoaossmpsS1.a1 |
Description: | At begining of period, the bypass blower on MAOS A failed. This was replaced by the instrument operator, but operator was unable to reprogram the PID flow controller. The blower ran approximately 50% faster than designed flow rate. It is unlikely this significantly affected the flow dynamics or transmission efficiencies in the MAOS A sampling stack. The data is colored transparent, but the DQR is filed to alert the user. No specific degradation of signals were noted at this time. |
Measurements: | maoaossmpsS1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/29/2014 | 0000 | 12/28/2015 | 0000 | Does not affect quality |
Subject: | MAO/SMPS/S1 - Reprocess: SMPS data |
DataStreams: | maoaossmpsS1.a1 |
Description: | No affect on data quality |
Measurements: | maoaossmpsS1.a1:
|