DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D121217.3 | TWP/MWR3C/C1 - Missing data | twpmwr3cC1.b1 | Missing |
D130416.2 | TWP/MWR3C/C1 - Intermittent bad data | twpmwr3cC1.b1 | Suspect |
D130416.3 | TWP/MWR3C/C1 - Incorrect data- bad calibration | twpmwr3cC1.b1 | Incorrect |
D131114.2 | TWP/MWR3C/C1 - Incorrect data | twpmwr3cC1.b1 | Incorrect |
D140115.1 | TWP/MWR3C/C1 - Instrument not calibrated | twpmwr3cC1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/08/2013 | 2100 | 06/07/2013 | 0100 | Missing |
03/17/2013 | 1100 | 03/25/2013 | 2200 | Missing |
06/12/2013 | 0500 | 07/12/2013 | 0600 | Missing |
10/13/2012 | 1300 | 10/17/2012 | 2200 | Missing |
12/16/2012 | 0100 | 02/27/2013 | 0200 | Missing |
Subject: | TWP/MWR3C/C1 - Missing data |
DataStreams: | twpmwr3cC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | twpmwr3cC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/20/2012 | 1100 | 07/19/2013 | 0000 | Suspect |
Subject: | TWP/MWR3C/C1 - Intermittent bad data |
DataStreams: | twpmwr3cC1.b1 |
Description: | The radiometer has intermittent communication issues and part of the brightness temperature data is missing. The 30-feet cable connecting the positioner to the main junction box was replaced on 7/19. |
Measurements: | twpmwr3cC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/27/2013 | 0200 | 02/28/2013 | 2359 | Incorrect |
Subject: | TWP/MWR3C/C1 - Incorrect data- bad calibration |
DataStreams: | twpmwr3cC1.b1 |
Description: | The radiometer was started but the calibration was incorrect. |
Measurements: | twpmwr3cC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/19/2013 | 0000 | 10/31/2013 | 2200 | Incorrect |
Subject: | TWP/MWR3C/C1 - Incorrect data |
DataStreams: | twpmwr3cC1.b1 |
Description: | The hardware configuration of the W-band receiver was corrupted and the data were saturated. The K-band readings were constantly zero. The vendor suspects a failure of the noise diode. On 10/9/2013 a spare K-band receiver was installed and on 10/31 the settings of the W-band receiver were restored. |
Measurements: | twpmwr3cC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/31/2013 | 2100 | 03/19/2014 | 1313 | Incorrect |
Subject: | TWP/MWR3C/C1 - Instrument not calibrated |
DataStreams: | twpmwr3cC1.b1 |
Description: | The radiometer is not calibrating. On February 3 the radiometer was moved to avoid obstructions in the field of view. However the settings of the mean radiating temperature had to be adjusted. |
Measurements: | twpmwr3cC1.b1:
|