DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160531.1 | ASI/AOSMET/M1 - Reprocessed and Corrected: Wind sensor not aligned to north | asiaosmetM1.a1 | Does not affect quality |
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaosmetM1.a1 | Incorrect |
D161013.5 | ASI/AOSMET/M1/M1 - Computer disk failure | asiaosmetM1.a1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/23/2016 | 0000 | 05/31/2016 | 1646 | Does not affect quality |
Subject: | ASI/AOSMET/M1 - Reprocessed and Corrected: Wind sensor not aligned to north |
DataStreams: | asiaosmetM1.a1 |
Description: | DATA REPROCESSED and CORRECTED: The wind sensor was not aligned to true north, resulting in wind direction data 16 degrees too high. The data were archived November 2016. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Red( Incorrect) REPROCESSED and CORRECTED DATA ARCHIVED:2016-11 REPROCESS IDENTIFICATION NUMBER: RID 1470 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The wind sensor was not aligned to true north, resulting in wind direction data 16 degrees too high. |
Measurements: | asiaosmetM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2016 | 0042 | 06/16/2016 | 1345 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
DataStreams: | asiaosmetM1.a1 |
Description: | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements: | asiaosmetM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2016 | 1500 | 10/13/2016 | 1800 | Does not affect quality |
Subject: | ASI/AOSMET/M1/M1 - Computer disk failure |
DataStreams: | asiaosmetM1.a1 |
Description: | Data are intermittently missing from 10/4 until the computer disk was replaced on 10/13. Data are completely missing 10/12 at 11:00 UTC until 10/13 at 18:00 UTC. |
Measurements: | asiaosmetM1.a1:
|