DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D170428.3 | ENA/HTDMA/NEPHELOMETER/C1 - Dry air supply failing | enaaoshtdmaC1.a1 | Suspect |
D170428.9 | ENA/HTDMA/NEPHELOMETER/C1 - Dry air supply high RH | enaaoshtdmaC1.a1 | Suspect |
D170519.7 | ENA/HTDMA/NEPHELOMETER/C1 - No dry air | enaaoshtdmaC1.a1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/25/2017 | 1100 | 04/26/2017 | 1200 | Suspect |
Subject: | ENA/HTDMA/NEPHELOMETER/C1 - Dry air supply failing |
DataStreams: | enaaoshtdmaC1.a1 |
Description: | Pentras dry air generator failing and offline for maintenance. Dry air supply to HT-DMA and humidigraph/nephelometers affected. Instruments working well, but with dry RH of around 30% instead <5%. |
Measurements: | enaaoshtdmaC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/26/2017 | 1100 | 04/28/2017 | 0810 | Suspect |
Subject: | ENA/HTDMA/NEPHELOMETER/C1 - Dry air supply high RH |
DataStreams: | enaaoshtdmaC1.a1 |
Description: | Humidigraph valve o-ring broken and valve stuck after o-ring replacement. Dry and humid air mixing. Instruments working well, except dry air RH different than nominal. HT-DMA affected, as well, as it is using the same dry air source. |
Measurements: | enaaoshtdmaC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/19/2017 | 0200 | 05/19/2017 | 1015 | Does not affect quality |
Subject: | ENA/HTDMA/NEPHELOMETER/C1 - No dry air |
DataStreams: | enaaoshtdmaC1.a1 |
Description: | Dry air supply not working. Data quality unaffected, instruments just working with RH different from the nominal. |
Measurements: | enaaoshtdmaC1.a1:
|