DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D161202.3 | ASI/NEPHELOMETER/M1 - Instrument offline | asiaosnephdry1mM1.b1 | Missing |
D170112.1 | ASI/NEPHELOMETER/M1 - Less efficient sample drying | asiaosnephdry1mM1.b1 | Does not affect quality |
D170203.7 | ASI/AOS/M1 - Mentor visit | asiaosnephdry1mM1.b1 | Suspect |
D170223.2 | ASI/NEPHELOMETER/M1 - Nephelometer bulb replaced | asiaosnephdry1mM1.b1 | Incorrect |
D170406.1 | ASI/AETH/NEPHELOMETER/PSAP/M1 - Decreased impactor flow | asiaosnephdry1mM1.b1 | Does not affect quality |
D170410.1 | ASI/NEPHELOMETER/M1 - Lamp burned out | asiaosnephdry1mM1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/30/2016 | 1600 | 11/30/2016 | 1910 | Missing |
12/01/2016 | 1400 | 12/01/2016 | 1815 | Missing |
Subject: | ASI/NEPHELOMETER/M1 - Instrument offline |
DataStreams: | asiaosnephdry1mM1.b1 |
Description: | Instrument offline for hard drive backup. No data being recorded. |
Measurements: | asiaosnephdry1mM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/20/2016 | 1900 | 12/23/2016 | 1200 | Does not affect quality |
Subject: | ASI/NEPHELOMETER/M1 - Less efficient sample drying |
DataStreams: | asiaosnephdry1mM1.b1 |
Description: | Humidigraph not getting the nominal amount of dry air. Sample unaffected and just the initial drying is less effective. All actual RH readings are properly recorded in the data. |
Measurements: | asiaosnephdry1mM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/19/2017 | 1000 | 01/27/2017 | 1900 | Suspect |
Subject: | ASI/AOS/M1 - Mentor visit |
DataStreams: | asiaosnephdry1mM1.b1 |
Description: | Mentors at the site working on different instruments. Data quality occasionally affected. |
Measurements: | asiaosnephdry1mM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/23/2017 | 1532 | 02/23/2017 | 1553 | Incorrect |
Subject: | ASI/NEPHELOMETER/M1 - Nephelometer bulb replaced |
DataStreams: | asiaosnephdry1mM1.b1 |
Description: | Dry nephelometer bulb was replaced. No valid data from this short period. |
Measurements: | asiaosnephdry1mM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/05/2017 | 1500 | 04/06/2017 | 1230 | Does not affect quality |
Subject: | ASI/AETH/NEPHELOMETER/PSAP/M1 - Decreased impactor flow |
DataStreams: | asiaosnephdry1mM1.b1 |
Description: | Impactor flow below nominal due to aethalometer failure. Impactor cut-point affected, nephelometer and PSAP data otherwise not affected. |
Measurements: | asiaosnephdry1mM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/01/2017 | 0625 | 04/01/2017 | 1149 | Incorrect |
Subject: | ASI/NEPHELOMETER/M1 - Lamp burned out |
DataStreams: | asiaosnephdry1mM1.b1 |
Description: | Nephelometer lamp burned out. No valid data from this time period. |
Measurements: | asiaosnephdry1mM1.b1:
|