Data Quality Reports for Session: 186091 User: rochettesb1 Completed: 06/12/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D130718.6FKB/TSI/M1 - Data processing problemfkbtsiskycoverM1.b1Does not affect quality
D170301.10FKB/TSI/M1 - Reprocess: Global attribute units incorrect for cam_mir_distfkbtsiskycoverM1.b1Incorrect


DQRID : D130718.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/14/2007000001/01/20082359Does not affect quality
Subject:
FKB/TSI/M1 - Data processing problem
DataStreams:fkbtsiskycoverM1.b1
Description:
Values of count_opaque and count_thin are reals but should be integers (units = pixels).
Measurements:fkbtsiskycoverM1.b1:
  • count_opaque
  • count_thin


Back To Table of Contents

DQRID : D170301.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/14/2007000001/01/20082359Incorrect
Subject:
FKB/TSI/M1 - Reprocess: Global attribute units incorrect for cam_mir_dist
DataStreams:fkbtsiskycoverM1.b1
Description:
The global attribute cam_mir_dist was converted incorrectly. The raw 
camera.mirror.distance, in inches, was multiplied by 10 and assigned a unit of mm.
Measurements:fkbtsiskycoverM1.b1:
  • base_time


Back To Table of Contents



END OF DATA