DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160531.1 | ASI/AOSMET/M1 - Reprocessed and Corrected: Wind sensor not aligned to north | asiaosmetM1.a1 | Does not affect quality |
D160607.3 | ASI/SONDE/S1 - Reprocessed and Corrected: Incorrect surface RH value | asisondewnpnS1.b1 | Does not affect quality |
D160607.4 | ASI/SONDE/S1 - Reprocessed and Corrected: Incorrect station altitude | asisondewnpnS1.b1 | Does not affect quality |
D160608.3 | ASI/TRACEGAS/M1 - Air leak | asiaoscoM1.b1 | Incorrect |
D160616.5 | ASI/CCN/M1 - Reprocessed and Corrected: Duplicate timestamps | asiaosccn200M1.a1 | Does not affect quality |
D160617.14 | ASI/AOS/CCN/HTDMA/NEPHELOMETER/UHSAS/M1 - Site installation | asiaosccn200M1.a1, asiaosimpactorM1.b1, asiaosnephdryM1.b1, asiaosuhsasM1.a1 | Incorrect |
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaoscpcfM1.b1, asiaosuhsasM1.a1, asiaosmetM1.a1, asiaosimpactorM1.b1, asiaosnephdryM1.b1, asiaoscoM1.b1, asiaosccn200M1.a1 | Incorrect |
D160622.1 | ASI/OZONE/PTR-MS/TRACEGAS/M1 - Sampling cabin air | asiaosso2M1.b1 | Incorrect |
D160707.5 | ASI/NEPHELOMETER/M1 - Instrument | asiaosnephdryM1.b1 | Missing |
D160708.10 | ASI/UHSAS/M1 - Instrument offline | asiaosuhsasM1.a1 | Missing |
D160708.3 | ASI/AOS/M1 - Impactor state not read | asiaosimpactorM1.b1 | Does not affect quality |
D160708.6 | ASI/AOS/M1 - Incorrect readout of Impactor state in 10-um setting not working | asiaosimpactorM1.b1 | Incorrect |
D160708.8 | ASI/SONDE/S1 - Reprocessed and Corrected: Incorrect sounding number | asisondewnpnS1.b1 | Does not affect quality |
D160713.1 | ASI/SONDE/S1 - RH suspect, possible sensor icing | asisondewnpnS1.b1 | Incorrect |
D160801.4 | ASI/SONDE/S1 - Reprocessed and Corrected: Incorrect lat and lon | asisondewnpnS1.b1 | Does not affect quality |
D160822.4 | ASI/CCN/M1 - Incorrect flows | asiaosccn200M1.a1 | Incorrect |
D160824.10 | ASI/UHSAS/M1 - Unstable flows | asiaosuhsasM1.a1 | Incorrect |
D160907.12 | ASI/CPC/M1 - Instrument testing | asiaoscpcfM1.b1 | Incorrect |
D160907.13 | ASI/CPC/M1 - Instrument maintence | asiaoscpcfM1.b1 | Incorrect |
D161013.5 | ASI/AOSMET/M1/M1 - Computer disk failure | asiaosmetM1.a1 | Does not affect quality |
D161024.2 | ASI/CPC/M1 - Collection | asiaoscpcfM1.b1 | Does not affect quality |
D161109.4 | ASI/TRACEGAS/M1 - Sampling cabin air. | asiaosso2M1.b1, asiaoscoM1.b1 | Incorrect |
D161116.1 | ASI/CCN/M1 - Concentration mismatch | asiaosccn200M1.a1 | Suspect |
D161116.5 | ASI/CCN/M1 - Instrument troubleshooting | asiaosccn200M1.a1 | Incorrect |
D161202.3 | ASI/NEPHELOMETER/M1 - Instrument offline | asiaosnephdryM1.b1 | Missing |
D161202.8 | ASI/CPC/M1 - Counts periodically drifting to zero | asiaoscpcfM1.b1 | Incorrect |
D161206.3 | ASI/CPC/M1 - Saturator wick changed out | asiaoscpcfM1.b1 | Incorrect |
D161209.7 | ASI/SONDE/S1 - Missing launches | asisondewnpnS1.b1 | Missing |
D161216.9 | ASI/CPC/M1 - Debris found in inner nozzle | asiaoscpcfM1.b1 | Incorrect |
D170109.18 | ASI/SONDE/S1 - Suspect values at start of launch | asisondewnpnS1.b1 | Incorrect |
D170112.1 | ASI/NEPHELOMETER/M1 - Less efficient sample drying | asiaosnephdryM1.b1 | Does not affect quality |
D170202.12 | ASI/CPC/SMPS/M1 - ACSM Calibration | asiaoscpcfM1.b1 | Missing |
D170202.13 | ASI/CPC/M1 - Gillibrator Flow Rate Test | asiaoscpcfM1.b1 | Incorrect |
D170202.20 | ASI/CPC/M1 - CPCF/SMPS comparison | asiaoscpcfM1.b1 | Suspect |
D170202.5 | ASI/CPC/M1 - Adding quick connect to vacuum line | asiaoscpcfM1.b1 | Incorrect |
D170202.7 | ASI/CPC/UHSAS/M1 - HEPA filter testing | asiaoscpcfM1.b1, asiaosuhsasM1.a1 | Incorrect |
D170202.8 | ASI/CPC/M1 - Cleaning & Calibration | asiaoscpcfM1.b1 | Missing |
D170203.7 | ASI/AOS/M1 - Mentor visit | asiaosccn200M1.a1, asiaosimpactorM1.b1, asiaosnephdryM1.b1, asiaosuhsasM1.a1 | Suspect |
D170208.3 | ASI/OZONE/TRACEGAS/M1 - Trace gas inlet blower failure | asiaosso2M1.b1, asiaoscoM1.b1 | Incorrect |
D170208.4 | ASI/OZONE/TRACEGAS/M1 - Instrument computer failure | asiaosso2M1.b1, asiaoscoM1.b1 | Missing |
D170210.5 | ASI/CPC/M1 - Instrument Butanol fill not initializing, cpc ran dry | asiaoscpcfM1.b1 | Incorrect |
D170210.6 | ASI/CPC/M1 - Instrument Data Missing | asiaoscpcfM1.b1 | Incorrect |
D170213.4 | ASI/CPC/M1 - Debris Blocking CPC Optics Block | asiaoscpcfM1.b1 | Incorrect |
D170214.4 | ASI/CPC/M1 - Instrument maintenance | asiaoscpcfM1.b1 | Missing |
D170223.2 | ASI/NEPHELOMETER/M1 - Nephelometer bulb replaced | asiaosnephdryM1.b1 | Incorrect |
D170406.1 | ASI/AETH/NEPHELOMETER/PSAP/M1 - Decreased impactor flow | asiaosnephdryM1.b1 | Does not affect quality |
D170410.1 | ASI/NEPHELOMETER/M1 - Lamp burned out | asiaosnephdryM1.b1 | Incorrect |
D170421.5 | ASI/CPC/M1 - Troubleshooting and replacing failed thermistor | asiaoscpcfM1.b1 | Missing |
D170426.2 | ASI/OZONE/TRACEGAS/M1 - Inlet blower replacement | asiaosso2M1.b1, asiaoscoM1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/23/2016 | 0000 | 05/31/2016 | 1646 | Does not affect quality |
Subject: | ASI/AOSMET/M1 - Reprocessed and Corrected: Wind sensor not aligned to north |
DataStreams: | asiaosmetM1.a1 |
Description: | DATA REPROCESSED and CORRECTED: The wind sensor was not aligned to true north, resulting in wind direction data 16 degrees too high. The data were archived November 2016. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Red( Incorrect) REPROCESSED and CORRECTED DATA ARCHIVED:2016-11 REPROCESS IDENTIFICATION NUMBER: RID 1470 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The wind sensor was not aligned to true north, resulting in wind direction data 16 degrees too high. |
Measurements: | asiaosmetM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/05/2016 | 0533 | 06/05/2016 | 0700 | Does not affect quality |
Subject: | ASI/SONDE/S1 - Reprocessed and Corrected: Incorrect surface RH value |
DataStreams: | asisondewnpnS1.b1 |
Description: | DATA REPROCESSED and CORRECTED: Site operations input the incorrect site altitude data in the DigiCORA software during deployment. 84.7m was input. 75.9m is the correct value. The data were archived September 2016. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2016-09 REPROCESS IDENTIFICATION NUMBER: RID 1473 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- An incorrect surface RH value was entered for this sounding. |
Measurements: | asisondewnpnS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/30/2016 | 1124 | 06/07/2016 | 2044 | Does not affect quality |
Subject: | ASI/SONDE/S1 - Reprocessed and Corrected: Incorrect station altitude |
DataStreams: | asisondewnpnS1.b1 |
Description: | DATA REPROCESSED and CORRECTED: Site operations input the incorrect site altitude data in the DigiCORA software during deployment. 84.7m was input. 75.9m is the correct value. The data were archived September 2016. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Yellow (Suspect) REPROCESSED and CORRECTED DATA ARCHIVED:2016-09 REPROCESS IDENTIFICATION NUMBER: RID 1474 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- Site operations input the incorrect site altitude data in the DigiCORA software during deployment. 84.7m was input. 75.9m is the correct value. |
Measurements: | asisondewnpnS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/03/2016 | 1549 | 06/06/2016 | 1958 | Incorrect |
Subject: | ASI/TRACEGAS/M1 - Air leak |
DataStreams: | asiaoscoM1.b1 |
Description: | Filter change resulted in total leak to cabin air. Data irrecoverable. |
Measurements: | asiaoscoM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/17/2016 | 1920 | 05/31/2016 | 1847 | Does not affect quality |
Subject: | ASI/CCN/M1 - Reprocessed and Corrected: Duplicate timestamps |
DataStreams: | asiaosccn200M1.a1 |
Description: | DATA REPROCESSED and CORRECTED: Data saved by the instrument had duplicate time stamps due to the instrument not properly handling fractional seconds. The raw data has been corrected and reprocessed. The data quality is not affected. These data were Archived in July 2016. Please retrieve the reprocessed data for use in your analysis. You may find these data using the ARM Discovery tool at the following link: http://www.archive.arm.gov/discovery/#v/results/s/s::aosccn200/fsite::asi/ffac::asi.M1 ORIGINAL DQR CATEGORY: Transparent(Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2016-07 REPROCESS IDENTIFICATION NUMBER: RID 1476 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- Data saved by the instrument had duplicate timestamps due to the instrument not properly handling fractional seconds. Data will be reprocessed to correct for this issue. Data quality not affected. |
Measurements: | asiaosccn200M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/22/2016 | 0000 | 05/31/2016 | 2359 | Incorrect |
Subject: | ASI/AOS/CCN/HTDMA/NEPHELOMETER/UHSAS/M1 - Site installation |
DataStreams: | asiaosccn200M1.a1, asiaosimpactorM1.b1, asiaosnephdryM1.b1, asiaosuhsasM1.a1 |
Description: | Site installation, mentors on site. No valid data. |
Measurements: | asiaosccn200M1.a1:
asiaosimpactorM1.b1:
asiaosnephdryM1.b1:
asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2016 | 0042 | 06/16/2016 | 1345 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
DataStreams: | asiaoscpcfM1.b1, asiaosuhsasM1.a1, asiaosmetM1.a1, asiaosimpactorM1.b1, asiaosnephdryM1.b1, asiaoscoM1.b1, asiaosccn200M1.a1 |
Description: | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements: | asiaosccn200M1.a1:
asiaoscoM1.b1:
asiaosmetM1.a1:
asiaosimpactorM1.b1:
asiaosnephdryM1.b1:
asiaosuhsasM1.a1:
asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/25/2016 | 0837 | 05/25/2016 | 1251 | Incorrect |
Subject: | ASI/OZONE/PTR-MS/TRACEGAS/M1 - Sampling cabin air |
DataStreams: | asiaosso2M1.b1 |
Description: | Inlet system turned off. Sampling cabin air. No effect on NO/NOx/NOy. |
Measurements: | asiaosso2M1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/24/2016 | 1824 | 06/26/2016 | 2322 | Missing |
Subject: | ASI/NEPHELOMETER/M1 - Instrument |
DataStreams: | asiaosnephdryM1.b1 |
Description: | Data missing due to an issue with communications between the instrument and the PC. |
Measurements: | asiaosnephdryM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/08/2016 | 1435 | 07/08/2016 | 1615 | Missing |
Subject: | ASI/UHSAS/M1 - Instrument offline |
DataStreams: | asiaosuhsasM1.a1 |
Description: | Instrument offline for flow adjustment. |
Measurements: | asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/23/2016 | 1300 | 05/23/2016 | 1505 | Does not affect quality |
Subject: | ASI/AOS/M1 - Impactor state not read |
DataStreams: | asiaosimpactorM1.b1 |
Description: | Impactor state was not read back from the limit switches. Impactor still switching and the impactor state data is valid. This applies only to raw data as the read-back state is only in raw data. |
Measurements: | asiaosimpactorM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/21/2016 | 1100 | 07/20/2016 | 1807 | Incorrect |
Subject: | ASI/AOS/M1 - Incorrect readout of Impactor state in 10-um setting not working |
DataStreams: | asiaosimpactorM1.b1 |
Description: | The limit switch in the 10-um position was not working. Operator found a loose wire and resoldered. |
Measurements: | asiaosimpactorM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/30/2016 | 2329 | 07/01/2016 | 0710 | Does not affect quality |
Subject: | ASI/SONDE/S1 - Reprocessed and Corrected: Incorrect sounding number |
DataStreams: | asisondewnpnS1.b1 |
Description: | DATA REPROCESSED and CORRECTED: The data quality is not affected. Metadata parameter "Sounding" indicating the sounding number was input incorrectly with month number "6" used instead of month number "7". 06302016-2329GMT launch had Sounding = 900601161; it is now 900701161. 07012016-0523GMT launch had Sounding = 900601162; it is now 900701162. The data were archived in March 2017. Please retrieve the reprocessed data for use in your analysis. You may find these data using the ARM Discovery tool at the following link: http://www.archive.arm.gov/discovery/#v/results/s/finst::{inst}/fsite::{site}/ffac::{site}.{fac} ORIGINAL DQR CATEGORY: TRansparent( Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-03 REPROCESS IDENTIFICATION NUMBER: RID 1483 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The data quality is not affected. Metedata parameter "Sounding" indicating the sounding number was input incorrectly with month number "6" used instead of month number "7". 06302016-2329GMT launch has Sounding = 900601161; it should be 900701161 07012016-0523GMT launch has Sounding = 900601162; it should be 900701162 |
Measurements: | asisondewnpnS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/11/2016 | 1145 | 07/11/2016 | 1218 | Incorrect |
Subject: | ASI/SONDE/S1 - RH suspect, possible sensor icing |
DataStreams: | asisondewnpnS1.b1 |
Description: | RH data during this time period is suspect due to possible sensor icing. |
Measurements: | asisondewnpnS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/01/2016 | 0229 | 08/01/2016 | 0323 | Does not affect quality |
Subject: | ASI/SONDE/S1 - Reprocessed and Corrected: Incorrect lat and lon |
DataStreams: | asisondewnpnS1.b1 |
Description: | DATA REPROCESSED and CORRECTED: The initial lat/lon settings were corrupted. Data has been reprocessed, collected, and ingested to reflect the correct lat/lon settings. The data were archived March 2017. Please retrieve the reprocessed data for use in your analysis. You may find these data using the ARM Discovery tool at the following link: http://www.archive.arm.gov/discovery/#v/results/s/finst::{inst}/fsite::{site}/ffac::{site}.{fac} ORIGINAL DQR CATEGORY: Red( Incorrect) REPROCESSED and CORRECTED DATA ARCHIVED:2017-03 REPROCESS IDENTIFICATION NUMBER: RID 1485 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- Somehow the initial Lat/Lon settings were corrupted. Data has been reprocessed and needs to be collected and ingested. |
Measurements: | asisondewnpnS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/20/2016 | 0200 | 08/20/2016 | 1530 | Incorrect |
Subject: | ASI/CCN/M1 - Incorrect flows |
DataStreams: | asiaosccn200M1.a1 |
Description: | Instrument flows incorrect due to water in the flow lines. |
Measurements: | asiaosccn200M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/22/2016 | 1600 | 08/22/2016 | 1900 | Incorrect |
Subject: | ASI/UHSAS/M1 - Unstable flows |
DataStreams: | asiaosuhsasM1.a1 |
Description: | Cleaning the internal flow lines and testing proper operation after a possible obstruction in the flow lines. No valid data from this period. |
Measurements: | asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/03/2016 | 1314 | 09/03/2016 | 1327 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument testing |
DataStreams: | asiaoscpcfM1.b1 |
Description: | During this time tests were being run with a HEPA filter on the inlet. Data is not representative of ambient aerosol concentrations |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1004 | 09/06/2016 | 1029 | Incorrect |
09/06/2016 | 1815 | 09/06/2016 | 1819 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument maintence |
DataStreams: | asiaoscpcfM1.b1 |
Description: | During these intervals, connections were being removed from the CPC to prepare for it to take the place of the SMPS CPC. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2016 | 1500 | 10/13/2016 | 1800 | Does not affect quality |
Subject: | ASI/AOSMET/M1/M1 - Computer disk failure |
DataStreams: | asiaosmetM1.a1 |
Description: | Data are intermittently missing from 10/4 until the computer disk was replaced on 10/13. Data are completely missing 10/12 at 11:00 UTC until 10/13 at 18:00 UTC. |
Measurements: | asiaosmetM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1030 | 09/06/2016 | 1815 | Does not affect quality |
Subject: | ASI/CPC/M1 - Collection |
DataStreams: | asiaoscpcfM1.b1 |
Description: | The CPC associated with the SMPS was sampling in place of the CPC associated with this CPCF measurement. The CPC's are the same model number |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/25/2016 | 2046 | 10/26/2016 | 1135 | Incorrect |
Subject: | ASI/TRACEGAS/M1 - Sampling cabin air. |
DataStreams: | asiaosso2M1.b1, asiaoscoM1.b1 |
Description: | Inlet blower turned off so instrument sampling cabin air. |
Measurements: | asiaosso2M1.b1:
asiaoscoM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/20/2016 | 1530 | 09/16/2016 | 1500 | Suspect |
Subject: | ASI/CCN/M1 - Concentration mismatch |
DataStreams: | asiaosccn200M1.a1 |
Description: | Concentration mismatch between the CCN columns at the same supersaturation set point after water was removed from one of the sheath air lines. Not clear which column was affected. |
Measurements: | asiaosccn200M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/16/2016 | 1500 | 11/16/2016 | 1700 | Incorrect |
Subject: | ASI/CCN/M1 - Instrument troubleshooting |
DataStreams: | asiaosccn200M1.a1 |
Description: | Troubleshooting instrument flow issues. No valid data. |
Measurements: | asiaosccn200M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/30/2016 | 1600 | 11/30/2016 | 1910 | Missing |
12/01/2016 | 1400 | 12/01/2016 | 1815 | Missing |
Subject: | ASI/NEPHELOMETER/M1 - Instrument offline |
DataStreams: | asiaosnephdryM1.b1 |
Description: | Instrument offline for hard drive backup. No data being recorded. |
Measurements: | asiaosnephdryM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/10/2016 | 2129 | 11/13/2016 | 1209 | Incorrect |
12/10/2016 | 0524 | 12/10/2016 | 1418 | Incorrect |
11/25/2016 | 1605 | 11/26/2016 | 0822 | Incorrect |
11/20/2016 | 1926 | 11/21/2016 | 1528 | Incorrect |
11/03/2016 | 1709 | 11/04/2016 | 1206 | Incorrect |
10/26/2016 | 1131 | 10/26/2016 | 1750 | Incorrect |
12/14/2016 | 0617 | 12/14/2016 | 1147 | Incorrect |
Subject: | ASI/CPC/M1 - Counts periodically drifting to zero |
DataStreams: | asiaoscpcfM1.b1 |
Description: | The CPC counts are periodically drifting to zero. The drift from concentrations comparable to the CPCU is somewhat abrupt, but not instantaneous. Placing a small pressure drop on the inlet (e.g. by measuring the flow rate) causes the counts to return to normal. The root cause of the problem has not yet been identified. More drifts to zero are possible in the future. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/21/2016 | 1740 | 11/21/2016 | 1818 | Incorrect |
Subject: | ASI/CPC/M1 - Saturator wick changed out |
DataStreams: | asiaoscpcfM1.b1 |
Description: | The saturator wick was changed out at this time. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/03/2016 | 1130 | 12/04/2016 | 1700 | Missing |
Subject: | ASI/SONDE/S1 - Missing launches |
DataStreams: | asisondewnpnS1.b1 |
Description: | Launches were not completed due to malfunction of GC25 equipment. |
Measurements: | asisondewnpnS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/15/2016 | 1455 | 12/15/2016 | 1706 | Incorrect |
Subject: | ASI/CPC/M1 - Debris found in inner nozzle |
DataStreams: | asiaoscpcfM1.b1 |
Description: | During this time, the CPCF was taken offline to clean the inner nozzle. Debris was found in the inner nozzle. During this time period, the CPC was either sampling cabin air or no data is available. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/02/2017 | 1731 | 02/02/2017 | 1732 | Incorrect |
08/01/2016 | 0825 | 08/01/2016 | 0826 | Incorrect |
09/24/2016 | 0829 | 09/24/2016 | 0830 | Incorrect |
08/08/2016 | 1132 | 08/08/2016 | 1132 | Incorrect |
08/05/2016 | 0814 | 08/05/2016 | 0815 | Incorrect |
02/09/2017 | 0530 | 02/09/2017 | 0531 | Incorrect |
09/02/2016 | 0526 | 09/02/2016 | 0527 | Incorrect |
10/30/2016 | 1131 | 10/30/2016 | 1132 | Incorrect |
09/02/2016 | 1429 | 09/02/2016 | 1430 | Incorrect |
09/27/2016 | 1133 | 09/27/2016 | 1134 | Incorrect |
09/29/2016 | 1132 | 09/29/2016 | 1133 | Incorrect |
08/28/2016 | 0829 | 08/28/2016 | 0830 | Incorrect |
07/22/2016 | 0525 | 07/22/2016 | 0526 | Incorrect |
08/29/2016 | 0822 | 08/29/2016 | 0823 | Incorrect |
09/17/2016 | 1429 | 09/17/2016 | 1430 | Incorrect |
08/24/2016 | 1426 | 08/24/2016 | 1426 | Incorrect |
08/21/2016 | 1125 | 08/21/2016 | 1126 | Incorrect |
08/07/2016 | 1128 | 08/07/2016 | 1129 | Incorrect |
09/22/2016 | 1435 | 09/22/2016 | 1436 | Incorrect |
06/24/2016 | 0517 | 06/24/2016 | 0518 | Incorrect |
10/14/2016 | 0526 | 10/14/2016 | 0527 | Incorrect |
11/11/2016 | 0525 | 11/11/2016 | 0526 | Incorrect |
01/16/2017 | 0531 | 01/16/2017 | 0531 | Incorrect |
11/25/2016 | 0527 | 11/25/2016 | 0528 | Incorrect |
09/23/2016 | 1426 | 09/23/2016 | 1427 | Incorrect |
09/12/2016 | 1429 | 09/12/2016 | 1430 | Incorrect |
08/23/2016 | 1426 | 08/23/2016 | 1427 | Incorrect |
01/13/2017 | 0529 | 01/13/2017 | 0530 | Incorrect |
09/09/2016 | 1501 | 09/09/2016 | 1502 | Incorrect |
06/25/2016 | 0524 | 06/25/2016 | 0525 | Incorrect |
01/30/2017 | 0529 | 01/30/2017 | 0530 | Incorrect |
08/16/2016 | 1129 | 08/16/2016 | 1130 | Incorrect |
06/22/2016 | 0514 | 06/22/2016 | 0515 | Incorrect |
09/10/2016 | 0828 | 09/10/2016 | 0829 | Incorrect |
10/23/2016 | 1125 | 10/23/2016 | 1126 | Incorrect |
09/20/2016 | 1512 | 09/20/2016 | 1513 | Incorrect |
08/27/2016 | 1723 | 08/27/2016 | 1723 | Incorrect |
08/05/2016 | 0525 | 08/05/2016 | 0526 | Incorrect |
09/21/2016 | 1505 | 09/21/2016 | 1505 | Incorrect |
08/26/2016 | 0526 | 08/26/2016 | 0527 | Incorrect |
08/20/2016 | 1430 | 08/20/2016 | 1430 | Incorrect |
11/13/2016 | 1144 | 11/13/2016 | 1145 | Incorrect |
11/29/2016 | 0607 | 11/29/2016 | 0608 | Incorrect |
02/14/2017 | 0532 | 02/14/2017 | 0533 | Incorrect |
09/06/2016 | 1428 | 09/06/2016 | 1429 | Incorrect |
08/23/2016 | 0527 | 08/23/2016 | 0528 | Incorrect |
08/08/2016 | 1422 | 08/08/2016 | 1423 | Incorrect |
08/17/2016 | 0828 | 08/17/2016 | 0829 | Incorrect |
11/06/2016 | 1130 | 11/06/2016 | 1131 | Incorrect |
07/08/2016 | 0522 | 07/08/2016 | 0523 | Incorrect |
08/19/2016 | 0826 | 08/19/2016 | 0827 | Incorrect |
02/03/2017 | 0528 | 02/03/2017 | 0529 | Incorrect |
09/01/2016 | 1426 | 09/01/2016 | 1427 | Incorrect |
08/26/2016 | 1427 | 08/26/2016 | 1428 | Incorrect |
08/20/2016 | 1127 | 08/20/2016 | 1128 | Incorrect |
08/02/2016 | 1130 | 08/02/2016 | 1131 | Incorrect |
06/17/2016 | 0523 | 06/17/2016 | 0524 | Incorrect |
06/18/2016 | 0527 | 06/18/2016 | 0528 | Incorrect |
08/29/2016 | 0523 | 08/29/2016 | 0524 | Incorrect |
01/23/2017 | 0526 | 01/23/2017 | 0527 | Incorrect |
09/05/2016 | 1431 | 09/05/2016 | 1432 | Incorrect |
10/28/2016 | 0525 | 10/28/2016 | 0526 | Incorrect |
10/09/2016 | 1132 | 10/09/2016 | 1133 | Incorrect |
09/18/2016 | 0830 | 09/18/2016 | 0831 | Incorrect |
12/12/2016 | 0528 | 12/12/2016 | 0529 | Incorrect |
08/22/2016 | 1425 | 08/22/2016 | 1426 | Incorrect |
08/13/2016 | 1725 | 08/13/2016 | 1726 | Incorrect |
09/03/2016 | 1433 | 09/03/2016 | 1434 | Incorrect |
08/03/2016 | 0820 | 08/03/2016 | 0821 | Incorrect |
09/15/2016 | 1131 | 09/15/2016 | 1132 | Incorrect |
06/27/2016 | 0526 | 06/27/2016 | 0527 | Incorrect |
06/20/2016 | 0520 | 06/20/2016 | 0521 | Incorrect |
11/27/2016 | 1121 | 11/27/2016 | 1122 | Incorrect |
08/12/2016 | 1417 | 08/12/2016 | 1418 | Incorrect |
08/06/2016 | 1449 | 08/06/2016 | 1450 | Incorrect |
09/28/2016 | 1124 | 09/28/2016 | 1125 | Incorrect |
09/19/2016 | 1427 | 09/19/2016 | 1428 | Incorrect |
06/16/2016 | 0603 | 06/16/2016 | 0604 | Incorrect |
08/30/2016 | 1423 | 08/30/2016 | 1423 | Incorrect |
Subject: | ASI/SONDE/S1 - Suspect values at start of launch |
DataStreams: | asisondewnpnS1.b1 |
Description: | Relative humidity measurements and dew point values are suspect at the beginning of each flight listed. RH drops to zero/near zero for unspecified reasons |
Measurements: | asisondewnpnS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/20/2016 | 1900 | 12/23/2016 | 1200 | Does not affect quality |
Subject: | ASI/NEPHELOMETER/M1 - Less efficient sample drying |
DataStreams: | asiaosnephdryM1.b1 |
Description: | Humidigraph not getting the nominal amount of dry air. Sample unaffected and just the initial drying is less effective. All actual RH readings are properly recorded in the data. |
Measurements: | asiaosnephdryM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/24/2017 | 1335 | 01/24/2017 | 1644 | Missing |
Subject: | ASI/CPC/SMPS/M1 - ACSM Calibration |
DataStreams: | asiaoscpcfM1.b1 |
Description: | The SMPS and CPCU were used to calibrate the ACSM during this time |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/25/2017 | 1020 | 01/25/2017 | 1022 | Incorrect |
Subject: | ASI/CPC/M1 - Gillibrator Flow Rate Test |
DataStreams: | asiaoscpcfM1.b1 |
Description: | Gillibrator was measuring the flow rate. CPCF not sampling from the atmosphere. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/27/2017 | 1031 | 01/27/2017 | 1107 | Suspect |
Subject: | ASI/CPC/M1 - CPCF/SMPS comparison |
DataStreams: | asiaoscpcfM1.b1 |
Description: | The CPCF was moved from its normal place to co-sample with the SMPS for a test on the inter-comparability of the two instruments during this time. Concentrations do not seem to deviate in this sampling configuration |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/21/2017 | 1019 | 01/21/2017 | 1023 | Incorrect |
Subject: | ASI/CPC/M1 - Adding quick connect to vacuum line |
DataStreams: | asiaoscpcfM1.b1 |
Description: | Probable pressure fluctuations during this time |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/21/2017 | 1029 | 01/21/2017 | 1030 | Incorrect |
01/25/2017 | 1015 | 01/25/2017 | 1018 | Incorrect |
Subject: | ASI/CPC/UHSAS/M1 - HEPA filter testing |
DataStreams: | asiaoscpcfM1.b1, asiaosuhsasM1.a1 |
Description: | HEPA filter on the inlet to all three instruments during this time for zero checks. All instruments went to zero counts |
Measurements: | asiaosuhsasM1.a1:
asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/21/2017 | 1035 | 01/21/2017 | 1048 | Missing |
Subject: | ASI/CPC/M1 - Cleaning & Calibration |
DataStreams: | asiaoscpcfM1.b1 |
Description: | CPC was cleaned and evaluated during this time |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/19/2017 | 1000 | 01/27/2017 | 1900 | Suspect |
Subject: | ASI/AOS/M1 - Mentor visit |
DataStreams: | asiaosccn200M1.a1, asiaosimpactorM1.b1, asiaosnephdryM1.b1, asiaosuhsasM1.a1 |
Description: | Mentors at the site working on different instruments. Data quality occasionally affected. |
Measurements: | asiaosccn200M1.a1:
asiaosimpactorM1.b1:
asiaosnephdryM1.b1:
asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/21/2016 | 2046 | 11/22/2016 | 1300 | Incorrect |
Subject: | ASI/OZONE/TRACEGAS/M1 - Trace gas inlet blower failure |
DataStreams: | asiaosso2M1.b1, asiaoscoM1.b1 |
Description: | MAOS C Trace gas inlet had the blower die. Though there are numbers, they should be viewed as incorrect. It also affects the PTR-MS. |
Measurements: | asiaosso2M1.b1:
asiaoscoM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/14/2016 | 0157 | 12/14/2016 | 0951 | Missing |
Subject: | ASI/OZONE/TRACEGAS/M1 - Instrument computer failure |
DataStreams: | asiaosso2M1.b1, asiaoscoM1.b1 |
Description: | Unit three computer failed. No data recorded until computer rebuilt remotely. |
Measurements: | asiaosso2M1.b1:
asiaoscoM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/08/2017 | 2000 | 02/09/2017 | 2014 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument Butanol fill not initializing, cpc ran dry |
DataStreams: | asiaoscpcfM1.b1 |
Description: | Due to a problem with the ambient pressure sensor, the butanol fill cycle was deactivated. When the butanol did not turn on, eventually the CPC dried out and concentrations dropped. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/08/2017 | 1744 | 02/08/2017 | 1751 | Incorrect |
02/08/2017 | 1408 | 02/08/2017 | 1450 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument Data Missing |
DataStreams: | asiaoscpcfM1.b1 |
Description: | Troubleshooting potential problems with the butanol fill solenoid valve and transistors. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/13/2017 | 0350 | 02/13/2017 | 1538 | Incorrect |
Subject: | ASI/CPC/M1 - Debris Blocking CPC Optics Block |
DataStreams: | asiaoscpcfM1.b1 |
Description: | During this period there was debris accumulated in the optics block that was affecting aerosol counts. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/14/2017 | 1529 | 02/14/2017 | 1604 | Missing |
Subject: | ASI/CPC/M1 - Instrument maintenance |
DataStreams: | asiaoscpcfM1.b1 |
Description: | The CPCF was removed for examination of the saturator wick for discoloration. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/23/2017 | 1532 | 02/23/2017 | 1553 | Incorrect |
Subject: | ASI/NEPHELOMETER/M1 - Nephelometer bulb replaced |
DataStreams: | asiaosnephdryM1.b1 |
Description: | Dry nephelometer bulb was replaced. No valid data from this short period. |
Measurements: | asiaosnephdryM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/05/2017 | 1500 | 04/06/2017 | 1230 | Does not affect quality |
Subject: | ASI/AETH/NEPHELOMETER/PSAP/M1 - Decreased impactor flow |
DataStreams: | asiaosnephdryM1.b1 |
Description: | Impactor flow below nominal due to aethalometer failure. Impactor cut-point affected, nephelometer and PSAP data otherwise not affected. |
Measurements: | asiaosnephdryM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/01/2017 | 0625 | 04/01/2017 | 1149 | Incorrect |
Subject: | ASI/NEPHELOMETER/M1 - Lamp burned out |
DataStreams: | asiaosnephdryM1.b1 |
Description: | Nephelometer lamp burned out. No valid data from this time period. |
Measurements: | asiaosnephdryM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/17/2017 | 1510 | 04/17/2017 | 1602 | Missing |
04/14/2017 | 1559 | 04/14/2017 | 1759 | Missing |
04/14/2017 | 1359 | 04/14/2017 | 1416 | Missing |
Subject: | ASI/CPC/M1 - Troubleshooting and replacing failed thermistor |
DataStreams: | asiaoscpcfM1.b1 |
Description: | The cpcu was taken offline for troubleshooting and replacing the failed thermistor. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/24/2017 | 0403 | 02/24/2017 | 1528 | Incorrect |
Subject: | ASI/OZONE/TRACEGAS/M1 - Inlet blower replacement |
DataStreams: | asiaosso2M1.b1, asiaoscoM1.b1 |
Description: | Bypass blower off (and replaced) |
Measurements: | asiaosso2M1.b1:
asiaoscoM1.b1:
|