DQRID : D160909.20
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/08/2016003509/08/20161501Incorrect
Subject:
ENA/AOS/C1 - Flow problem affecting concentrations
DataStreams:enaaoscpcfC1.a1
Description:
Due to a flow problem with another instrument on the same sample line, the concentrations 
reported by the CPC are not representative of the atmospheric concentration of particles
Measurements:enaaoscpcfC1.a1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D161025.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/26/2016083010/01/20160100Suspect
Subject:
ENA/AOS/C1 - Instrument mentors working at the site
DataStreams:enaaoscpcfC1.a1
Description:
Instrument mentors working at the site. Sample lines and electrical connections being 
reconfigured. Instruments that had no work done on them may still be affected.
Measurements:enaaoscpcfC1.a1:
  • dilution_flow_setpoint
  • laser_current
  • nozzle_pressure
  • ambient_temp
  • fraction_full
  • ambient_pressure
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • liquid_level
  • dilution_valve_position
  • optics_temp
  • orifice_pressure
  • condensation_temp
  • saturation_temp
  • cabinet_temp
  • dilution_flow_reading


Back To Table of Contents

DQRID : D170303.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/20/2016000003/03/20170000Does not affect quality
Subject:
ENA/AOS/C1 - Dry air supply intermittently not working
DataStreams:enaaoscpcfC1.a1
Description:
Pentras dry air supply intermittently not working. Dry air missing for the humidigraph and 
the HT-DMA. Dry air supply for those instruments partially replaced by regular undried 
compressed air, lowering the delivery pressure. No direct effect on other instruments seen.
Measurements:enaaoscpcfC1.a1:
  • East longitude(lon)
  • dilution_flow_setpoint
  • laser_current
  • nozzle_pressure
  • time
  • fraction_full
  • ambient_pressure
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • time_offset
  • liquid_level
  • dilution_valve_position
  • optics_temp
  • orifice_pressure
  • condensation_temp
  • Altitude above mean sea level(alt)
  • saturation_temp
  • North latitude(lat)
  • base_time
  • cabinet_temp
  • dilution_flow_reading


Back To Table of Contents

DQRID : D170320.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/22/2017174002/22/20171958Incorrect
Subject:
ENA/CPC/C1 - Changes to upstream plumbing, may not have been sampling ambient.
DataStreams:enaaoscpcfC1.a1
Description:
During this time period, changes were being made to upstream plumbing (i.e. the impactor). 
 The cpcf may not have been receiving all sample flow from the ambient during this time 
period.
Measurements:enaaoscpcfC1.a1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)