DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D170123.1 | SGP/MET/E11/E13/E15/E21/E31/E32/E33/E34/E35/E36/E37/E38/E39/E40/E41/E9 - Ice storm affected winds | sgpmetE13.b1 | Incorrect |
D170201.1 | SGP/MET/E13 - Optical rain gauge data missing | sgpmetE13.b1 | Missing |
D170221.5 | SGP/MET/STAMP/All - Freezing Rain Event | sgpmetE13.b1 | Does not affect quality |
D170303.6 | SGP/MET/E13 - Optical rain gauge data parsing error | sgpmetE13.b1 | Incorrect |
D170310.16 | SGP/Central Facility - Power Outage | sgpmetE13.b1 | Does not affect quality |
D170327.3 | SGP/MET/E13 - Present weather detector data incorrect after power outage | sgpmetE13.b1 | Incorrect |
D170327.5 | SGP/MET/E13 - Optical rain gauge data incorrect after power outage | sgpmetE13.b1 | Incorrect |
D170420.1 | SGP/MET/E13 - Wind speed incorrect | sgpmetE13.b1 | Incorrect |
D170531.1 | SGP/MET/E13 - Optical rain gauge data incorrect after power outage | sgpmetE13.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/14/2017 | 0000 | 01/16/2017 | 2359 | Incorrect |
Subject: | SGP/MET/E11/E13/E15/E21/E31/E32/E33/E34/E35/E36/E37/E38/E39/E40/E41/E9 - Ice storm |
DataStreams: | sgpmetE13.b1 |
Description: | An ice storm impaired the wind sensor mechanics during this time. At most sites, winds flatlined for a large part of 1/15. Data from 1/14 - 1/16 may not be accurate. |
Measurements: | sgpmetE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/03/2016 | 1616 | 01/31/2017 | 2057 | Missing |
Subject: | SGP/MET/E13 - Optical rain gauge data missing |
DataStreams: | sgpmetE13.b1 |
Description: | No optical rain gauge data are available during this time. |
Measurements: | sgpmetE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/13/2017 | 1000 | 01/16/2017 | 1200 | Does not affect quality |
Subject: | SGP/MET/STAMP/All - Freezing Rain Event |
DataStreams: | sgpmetE13.b1 |
Description: | A freezing rain event potentially affected some of the STAMP and MET tipping bucket rain gauges. The rain accumulation amounts generally agree (2 exceptions noted below) but the timing and rate of precipitation may disagree at times between the STAMP and MET. Instruments at E33 and E39 were found to differ more than the expected uncertainty and will be covered in additional DQRs if a problem is found. |
Measurements: | sgpmetE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/01/2017 | 0000 | 02/22/2017 | 2144 | Incorrect |
Subject: | SGP/MET/E13 - Optical rain gauge data parsing error |
DataStreams: | sgpmetE13.b1 |
Description: | A data acquisition problem caused incorrect optical rain gauge data to be reported. |
Measurements: | sgpmetE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/05/2017 | 0023 | 03/09/2017 | 1430 | Does not affect quality |
Subject: | SGP/Central Facility - Power Outage |
DataStreams: | sgpmetE13.b1 |
Description: | Automated scripts show that these instruments were affected by a power outage at the central facility resulting in missing data. Most of the instruments came back online on March 6th but some took longer. |
Measurements: | sgpmetE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/06/2017 | 1951 | 03/24/2017 | 1822 | Incorrect |
Subject: | SGP/MET/E13 - Present weather detector data incorrect after power outage |
DataStreams: | sgpmetE13.b1 |
Description: | A power interruption caused the data logger to read incorrect data from the present weather detector. |
Measurements: | sgpmetE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/06/2017 | 1951 | 03/23/2017 | 2127 | Incorrect |
Subject: | SGP/MET/E13 - Optical rain gauge data incorrect after power outage |
DataStreams: | sgpmetE13.b1 |
Description: | A power interruption caused the data logger to read incorrect data from the optical rain gauge. |
Measurements: | sgpmetE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/06/2017 | 1910 | 04/17/2017 | 1541 | Incorrect |
Subject: | SGP/MET/E13 - Wind speed incorrect |
DataStreams: | sgpmetE13.b1 |
Description: | A broken wind speed signal wire caused incorrect data during this time. |
Measurements: | sgpmetE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/23/2017 | 2128 | 05/16/2017 | 1827 | Incorrect |
Subject: | SGP/MET/E13 - Optical rain gauge data incorrect after power outage |
DataStreams: | sgpmetE13.b1 |
Description: | A power interruption caused the data logger to read incorrect data from the optical rain gauge. |
Measurements: | sgpmetE13.b1:
|