Data Quality Reports for Session: 186146 User: gallof1 Completed: 06/14/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160909.20ENA/AOS/C1 - Flow problem affecting concentrationsenaaoscpcfC1.a1Incorrect
D161025.6ENA/AOS/C1 - Instrument mentors working at the siteenaaoscpcfC1.a1Suspect
D170303.12ENA/AOS/C1 - Dry air supply intermittently not workingenaaoscpcfC1.a1Does not affect quality


DQRID : D160909.20
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/08/2016003509/08/20161501Incorrect
Subject:
ENA/AOS/C1 - Flow problem affecting concentrations
DataStreams:enaaoscpcfC1.a1
Description:
Due to a flow problem with another instrument on the same sample line, the concentrations 
reported by the CPC are not representative of the atmospheric concentration of particles
Measurements:enaaoscpcfC1.a1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)


Back To Table of Contents

DQRID : D161025.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/26/2016083010/01/20160100Suspect
Subject:
ENA/AOS/C1 - Instrument mentors working at the site
DataStreams:enaaoscpcfC1.a1
Description:
Instrument mentors working at the site. Sample lines and electrical connections being 
reconfigured. Instruments that had no work done on them may still be affected.
Measurements:enaaoscpcfC1.a1:
  • dilution_flow_setpoint
  • laser_current
  • nozzle_pressure
  • ambient_temp
  • fraction_full
  • ambient_pressure
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • liquid_level
  • dilution_valve_position
  • optics_temp
  • orifice_pressure
  • condensation_temp
  • saturation_temp
  • cabinet_temp
  • dilution_flow_reading


Back To Table of Contents

DQRID : D170303.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/20/2016000003/03/20170000Does not affect quality
Subject:
ENA/AOS/C1 - Dry air supply intermittently not working
DataStreams:enaaoscpcfC1.a1
Description:
Pentras dry air supply intermittently not working. Dry air missing for the humidigraph and 
the HT-DMA. Dry air supply for those instruments partially replaced by regular undried 
compressed air, lowering the delivery pressure. No direct effect on other instruments seen.
Measurements:enaaoscpcfC1.a1:
  • East longitude(lon)
  • dilution_flow_setpoint
  • laser_current
  • nozzle_pressure
  • time
  • fraction_full
  • ambient_pressure
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • time_offset
  • liquid_level
  • dilution_valve_position
  • optics_temp
  • orifice_pressure
  • condensation_temp
  • Altitude above mean sea level(alt)
  • saturation_temp
  • North latitude(lat)
  • base_time
  • cabinet_temp
  • dilution_flow_reading


Back To Table of Contents



END OF DATA