DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D140320.2 | NSA/MWRP/C1 - Thermal instability | nsamwrpC1.b1 | Incorrect |
D140320.1 | NSA/MWRP/C1 - Questionable calibration | nsamwrpC1.b1 | Suspect |
D141027.2 | NSA/MWRP/C1 - calibration change | nsamwrpC1.b1 | Incorrect |
D141212.4 | NSA/MWRP/C1 - Reprocessing Notice (RID 1262) | nsamwrpC1.b1 | Does not affect quality |
D141212.3 | NSA/MWRP/C1 - missing data | nsamwrpC1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-02-12 | 00:00:00 | 2014-02-23 | 00:00:00 | Incorrect |
Subject: | NSA/MWRP/C1 - Thermal instability |
---|---|
Affected Datastreams: | nsamwrpC1.b1 |
Description | The brightness temperatures and retrievals appear affected by some thermal instability. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-02-07 | 20:25:00 | 2014-02-10 | 14:26:00 | Suspect |
Subject: | NSA/MWRP/C1 - Questionable calibration |
---|---|
Affected Datastreams: | nsamwrpC1.b1 |
Description | The K-band channels didn't calibrate well with liquid nitrogen. On 2/10 the coefficients were updated with tip-derived values |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-09-05 | 00:00:00 | 2014-09-29 | 14:02:59 | Incorrect |
Subject: | NSA/MWRP/C1 - calibration change |
---|---|
Affected Datastreams: | nsamwrpC1.b1 |
Description | The calibration coefficients have slightly drifted affecting the brightness temperatures and the retrievals. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-11-01 | 00:00:00 | 2014-12-12 | 18:15:00 | Does not affect quality |
Subject: | NSA/MWRP/C1 - Reprocessing Notice (RID 1262) |
---|---|
Affected Datastreams: | nsamwrpC1.b1 |
Description | See below for details of reprocessing that was completed on 20150309. Original DQR category: 4 (Red, Incorrect). ----------------------------------------------------------- The instrument appears to be unstable and the brightness temperatures are too warm. This affects the retrievals as well. The temperature dependence of the calibration coefficients has slightly changed. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-12-11 | 00:00:00 | 2014-12-12 | 15:48:33 | Missing |
2014-12-03 | 00:00:00 | 2014-12-07 | 00:10:00 | Missing |
2014-11-04 | 00:00:00 | 2014-11-07 | 14:10:53 | Missing |
Subject: | NSA/MWRP/C1 - missing data |
---|---|
Affected Datastreams: | nsamwrpC1.b1 |
Description | Data are missing |
Measurements |
|