DQR ID Subject Affected Data Streams Data Quality Metric
D050601.1 SGP/SIRS/E13 - Reprocessed and Corrected: incorrect calibrations sgpsirsE13.b1 Does not affect quality
D121003.1 SGP/SIRS/E13 - Instrument problem sgpsirsE13.b1 Suspect
D060630.21 SGP/SIRS/E13 - Reprocessed: Longwave Calibration error sgpsirsE13.b1 Does not affect quality
D110725.2 SGP/SIRS/E13 - NIP damaged by lightning strike sgpsirsE13.b1 Incorrect
 
DQR ID: D050601.1    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2005-05-25 14:09:00 2005-06-02 22:05:00 Does not affect quality
Subject: SGP/SIRS/E13 - Reprocessed and Corrected: incorrect calibrations
Affected Datastreams: sgpsirsE13.b1
Description DATA REPROCESSED and CORRECTED: After a long data gap on 05/25/2005, presumably a power outage, short_direct_normal (NIP) was showing incorrect value as well as incorrect calibration coefficients in the data logger program. The SIRS. E13 data logger was power cycled (6/2/2005 GMT) and the correct program loaded into the SIRS data logger and storage module. The data during this time range were reprocessed to apply the correct calibration coefficients for Short_direct_normal data. The data were archived in September 2016. Please retrieve the reprocessed data for use in your analysis. You may find these data using the ARM Discovery tool at the following link: http://www.archive.arm.gov/discovery/#v/results/s/finst::sirs/fsite::sgp/ffac::sgp.E1 ORIGINAL DQR CATEGORY: Red( Incorrect) REPROCESSED and CORRECTED DATA ARCHIVED:2016-09 REPROCESS IDENTIFICATION NUMBER: RID 318 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- After a long data gap on 5/25/2005, presumably a power outage, the following datastreams were showing incorrect values as well as incorrect calibration coefficients in the datalogger program: down_long_hemisp_shaded (downwelling PIR instrument) up_long_hemisp (upwelling PIR instrument) short_direct_normal (NIP instrument) The SIRS.E13 datalogger was power cycled (6/2/2005 2200GMT) and the correct program loaded into the SIRS datalogger and storage module. The data during this time range can be recalibrated as follows: First divide value by incorrectly applied cal coeff (first number) then multiply by correct cal coeff (second number) short_direct_normal 120.28 124.58 up_long_netir 0.2386 0.2595 down_long_netir 0.3052 0.2304 Reapply cal coeffs below using up_long_netIR(uV), inst_up_long_dome_temp(K), inst_up_long_dome_temp(K) using procedure described in SIRS instrument manual. up_long_hemisp k0 -18.3400 W/m^2\n, k1 0.2595 W/(m^2*uV)\n, k2 1.0380 unitless\n, k3 -2.7420 unitless\n, kr 0.0007044 K/uV\n, Reapply cal coeffs below using down_long_netIR(uV),inst_down_long_dome_temp (K),inst_down_long_dome_temp(K). down_long_hemisp_shaded k0 -11.6300 W/m^2\n, k1 0.2304 W/(m^2*uV)\n, k2 1.0190 unitless\n, k3 -2.4230 unitless\n, kr 0.0007044 K/uV\n,
Measurements
  • down_long_hemisp_shaded_max (Downwelling longwave hemispheric irradiance, shaded pyrgeometer, maxima)
  • down_long_netir (Downwelling longwave hemispheric net infrared)
  • down_long_hemisp_shaded_min (Downwelling longwave hemispheric irradiance, shaded pyrgeometer, minima)
  • inst_down_long_shaded_case_temp (Instantaneous downwelling pyrgeometer case thermistor temperature, shaded pyrgeometer)
  • down_long_hemisp_shaded (Downwelling longwave hemispheric irradiance, shaded pyrgeometer)
  • up_long_hemisp_std (Upwelling longwave hemispheric irradiance, pyrgeometer, standard deviation)
  • down_long_hemisp_shaded_std (Downwelling longwave hemispheric irradiance, shaded pyrgeometer, standard deviation)
  • up_long_hemisp (Upwelling longwave hemispheric irradiance, pyrgeometer)
  • short_direct_normal (Shortwave direct normal irradiance, pyrheliometer)
  • up_long_netir (Upwelling longwave hemispheric net infrared)
  • inst_up_long_dome_temp (Instantaneous upwelling pyrgeometer dome thermistor temperature, pyrgeometer)
  • inst_up_long_case_temp (Instantaneous upwelling pyrgeometer case thermistor temperature, pyrgeometer)
  • up_long_hemisp_min (Upwelling longwave hemispheric irradiance, pyrgeometer, minima)
  • short_direct_normal_min (Shortwave direct normal irradiance, pyrheliometer, minima)
  • short_direct_normal_max (Shortwave direct normal irradiance, pyrheliometer, maxima)
  • inst_down_long_shaded_dome_temp (Instantaneous downwelling pyrgeometer dome thermistor temperature, shaded pyrgeometer)
  • short_direct_normal_std (Shortwave direct normal irradiance, pyrheliometer, standard deviation)
  • up_long_hemisp_max (Upwelling longwave hemispheric irradiance, pyrgeometer, maxima)
 
DQR ID: D121003.1    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2012-05-13 00:00:00 2012-09-20 20:45:00 Suspect
Subject: SGP/SIRS/E13 - Instrument problem
Affected Datastreams: sgpsirsE13.b1
Description up_long_hemisp measurements are intermittant and remain so after extensive troubleshooting. This problem will likly remain open ended until site ops can replace the cable which serves the up_long_hemisp instrument (i.e. the Eppley PIR).
Measurements
  • up_short_hemisp (Upwelling shortwave hemispheric irradiance, pyranometer)
  • up_long_hemisp_std (Upwelling longwave hemispheric irradiance, pyrgeometer, standard deviation)
  • up_long_hemisp (Upwelling longwave hemispheric irradiance, pyrgeometer)
  • up_long_netir (Upwelling longwave hemispheric net infrared)
  • inst_up_long_dome_temp (Instantaneous upwelling pyrgeometer dome thermistor temperature, pyrgeometer)
  • inst_up_long_case_temp (Instantaneous upwelling pyrgeometer case thermistor temperature, pyrgeometer)
  • up_long_hemisp_min (Upwelling longwave hemispheric irradiance, pyrgeometer, minima)
  • up_long_hemisp_max (Upwelling longwave hemispheric irradiance, pyrgeometer, maxima)
 
DQR ID: D060630.21    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2003-12-11 19:35:00 2006-02-16 19:05:00 Does not affect quality
Subject: SGP/SIRS/E13 - Reprocessed: Longwave Calibration error
Affected Datastreams: sgpsirsE13.b1
Description Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006.
Measurements
  • down_long_hemisp_shaded (Downwelling longwave hemispheric irradiance, shaded pyrgeometer)
  • down_long_netir (Downwelling longwave hemispheric net infrared)
  • up_long_netir (Upwelling longwave hemispheric net infrared)
  • up_long_hemisp (Upwelling longwave hemispheric irradiance, pyrgeometer)
 
DQR ID: D110725.2    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2011-06-12 05:39:00 2011-06-14 15:56:00 Incorrect
Subject: SGP/SIRS/E13 - NIP damaged by lightning strike
Affected Datastreams: sgpsirsE13.b1
Description Apparent lightning strike in the instrument cluster area resulted in a failed NIP. The SIRS was due for calibration change outs so the US, DS, DD and NIP were replaced. The updated program was loaded at approximately 1730 GMT. An estimate of direct normal values can be derived using the following formula: short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle)
Measurements
  • short_direct_normal (Shortwave direct normal irradiance, pyrheliometer)
  • short_direct_normal_min (Shortwave direct normal irradiance, pyrheliometer, minima)
  • short_direct_normal_max (Shortwave direct normal irradiance, pyrheliometer, maxima)
  • short_direct_normal_std (Shortwave direct normal irradiance, pyrheliometer, standard deviation)