DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D170726.4 | ASI/TRACEGAS/M1 - | asiaoscoM1.a1 | Incorrect |
D160622.1 | ASI/OZONE/PTR-MS/TRACEGAS/M1 - Sampling cabin air | asiaoscoM1.a1 | Incorrect |
D160608.3 | ASI/TRACEGAS/M1 - Air leak | asiaoscoM1.b1 | Incorrect |
D170726.9 | ASI/CO-ANALYZER/M1 - Reprocess: valve position fields incorrect | asiaoscoM1.a1 | Incorrect |
D170426.2 | ASI/OZONE/TRACEGAS/M1 - Inlet blower replacement | asiaosso2M1.01 | Incorrect |
D170208.3 | ASI/OZONE/TRACEGAS/M1 - Trace gas inlet blower failure | asiaoscoM1.01 | Incorrect |
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaossmpstrhM1.a1 | Incorrect |
D170208.4 | ASI/OZONE/TRACEGAS/M1 - Instrument computer failure | asiaoscoM1.01 | Missing |
D161109.4 | ASI/TRACEGAS/M1 - Sampling cabin air. | asiaoscoM1.a1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-04-03 | 16:53:33 | 2017-04-05 | 11:40:03 | Incorrect |
Subject: | ASI/TRACEGAS/M1 - |
---|---|
Affected Datastreams: | asiaoscoM1.a1 |
Description | Pump stopped. No useful data in this period |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-05-25 | 08:37:42 | 2016-05-25 | 12:51:35 | Incorrect |
Subject: | ASI/OZONE/PTR-MS/TRACEGAS/M1 - Sampling cabin air |
---|---|
Affected Datastreams: | asiaoscoM1.a1 |
Description | Inlet system turned off. Sampling cabin air. No effect on NO/NOx/NOy. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-06-03 | 15:49:13 | 2016-06-06 | 19:58:34 | Incorrect |
Subject: | ASI/TRACEGAS/M1 - Air leak |
---|---|
Affected Datastreams: | asiaoscoM1.b1 |
Description | Filter change resulted in total leak to cabin air. Data irrecoverable. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-06-13 | 18:00:00 | 2017-07-26 | 23:59:59 | Incorrect |
Subject: | ASI/CO-ANALYZER/M1 - Reprocess: valve position fields incorrect |
---|---|
Affected Datastreams: | asiaoscoM1.a1 |
Description | A bug in the a1- and b1-level AOS CO ingest processes caused the valve position fields to be incorrect. This had the trickle-down effect of flagging the primary fields as incorrect via embedded QC in the b1-level product. Reprocessing is being requested now that this ingest bug has been corrected. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-02-24 | 04:03:06 | 2017-02-24 | 15:28:46 | Incorrect |
Subject: | ASI/OZONE/TRACEGAS/M1 - Inlet blower replacement |
---|---|
Affected Datastreams: | asiaosso2M1.01 |
Description | Bypass blower was off and has now been replaced. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-11-21 | 20:46:16 | 2016-11-22 | 13:00:49 | Incorrect |
Subject: | ASI/OZONE/TRACEGAS/M1 - Trace gas inlet blower failure |
---|---|
Affected Datastreams: | asiaoscoM1.01 |
Description | MAOS C Trace gas inlet had the blower die. Though there are numbers, they should be viewed as incorrect. It also affects the PTR-MS. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-06-16 | 00:42:00 | 2016-06-16 | 13:45:00 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
---|---|
Affected Datastreams: | asiaossmpstrhM1.a1 |
Description | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-12-14 | 01:57:16 | 2016-12-14 | 09:51:51 | Missing |
Subject: | ASI/OZONE/TRACEGAS/M1 - Instrument computer failure |
---|---|
Affected Datastreams: | asiaoscoM1.01 |
Description | Unit three computer failed. No data recorded until computer rebuilt remotely. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-10-25 | 20:46:13 | 2016-10-26 | 11:35:27 | Incorrect |
Subject: | ASI/TRACEGAS/M1 - Sampling cabin air. |
---|---|
Affected Datastreams: | asiaoscoM1.a1 |
Description | Inlet blower turned off so instrument sampling cabin air. |
Measurements |
|