Data Quality Report for Order ID: 186966
User ID: pzuidema    Date completed: 2017-07-29
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D161109.2 ASI/OZONE/M1 - Not sampling outside air asiaoso3M1.a1 Incorrect
D170426.2 ASI/OZONE/TRACEGAS/M1 - Inlet blower replacement asiaoso3M1.a1 Incorrect
D160621.4 ASI/AOS/M1 - Manifold blower not working asiaosccn2colbavgM1.b1 Incorrect
D170208.4 ASI/OZONE/TRACEGAS/M1 - Instrument computer failure asiaosso2M1.a1 Missing
D170104.3 ASI/OZONE/M1 - Reprocessed and Corrected: asiaosozoneM1.a1 to asiaoso3M1.a1 asiaoso3M1.a1 Note
D170208.3 ASI/OZONE/TRACEGAS/M1 - Trace gas inlet blower failure asiaoso3M1.a1 Incorrect
 
DQR ID: D161109.2    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-10-25 20:46:13 2016-10-26 11:35:27 Incorrect
Subject: ASI/OZONE/M1 - Not sampling outside air
Affected Datastreams: asiaoso3M1.a1
Description Inlet turned off (or the blower turned off). So sampling was not on outside air.
Measurements
  • pressure_compensation_state (Pressure compensation state)
  • intensity_b (Intensity in cell B)
  • o3_coefficient (Instrument ozone coefficient)
  • lamp_level (Lamp level)
  • calibration_level_2 (Calibration ratio level 2)
  • flow_b (Flow in cell B)
  • diagnostic_voltage_mib_24 (Diagnostic +24 volts at measurement interface board)
  • lamp_voltage_bench (Lamp bench voltage)
  • diagnostic_voltage_mib_5 (Diagnostic +5 volts at measurement interface board)
  • o3_flags (Ozone flag string)
  • diagnostic_voltage_mb_3p3 (Diagnostic +3.3 volt at motherboard)
  • diagnostic_voltage_mib_15 (Diagnostic +15 volts at measurement interface board)
  • calibration_level_4 (Calibration ratio level 4)
  • gas_state (Gas state)
  • calibration_level_5 (Calibration ratio level 5)
  • o3_lamp_state (Ozone lamp state)
  • lamp_temperature (Lamp temperature)
  • noise_a (Electric noise in cell A)
  • diagnostic_voltage_mib_3p3 (Diagnostic +3.3 volts at measurement interface board)
  • o3_lamp_temperature (Ozone lamp temperature)
  • averaging_time (Instrument averaging time)
  • flow_a (Flow in cell A)
  • intensity_a (Intensity in cell A)
  • lamp_voltage_ozonizer (Lamp voltage ozonizer)
  • o3_background (Ozone background)
  • calibration_level_1 (Calibration ratio level 1)
  • range (Instrument range setting)
  • diagnostic_voltage_mb_15 (Diagnostic +15 volts at motherboard)
  • diagnostic_voltage_mb_5 (Diagnostic +5 volts at motherboard)
  • o3 (Ozone concentration)
  • diagnostic_voltage_mb_minus_3p3 (Diagnostic -3.3 volts at motherboard)
  • o3_pressure (Ozone pressure)
  • temperature_compensation_state (Temperature compensation state)
  • o3_bench_temperature (Bench temperature)
  • noise_b (Electric noise in cell B)
  • diagnostic_voltage_mib_minus_15 (Diagnostic -15 volts at measurement interface board)
  • diagnostic_voltage_mb_24 (Diagnostic +24 volts at motherboard)
  • calibration_level_3 (Calibration ratio level 3)
 
DQR ID: D170426.2    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2017-02-24 04:03:06 2017-02-24 15:28:46 Incorrect
Subject: ASI/OZONE/TRACEGAS/M1 - Inlet blower replacement
Affected Datastreams: asiaoso3M1.a1
Description Bypass blower was off and has now been replaced.
Measurements
  • pressure_compensation_state (Pressure compensation state)
  • intensity_b (Intensity in cell B)
  • o3_coefficient (Instrument ozone coefficient)
  • lamp_level (Lamp level)
  • calibration_level_2 (Calibration ratio level 2)
  • diagnostic_voltage_mib_24 (Diagnostic +24 volts at measurement interface board)
  • flow_b (Flow in cell B)
  • lamp_voltage_bench (Lamp bench voltage)
  • diagnostic_voltage_mib_5 (Diagnostic +5 volts at measurement interface board)
  • o3_flags (Ozone flag string)
  • diagnostic_voltage_mb_3p3 (Diagnostic +3.3 volt at motherboard)
  • diagnostic_voltage_mib_15 (Diagnostic +15 volts at measurement interface board)
  • calibration_level_4 (Calibration ratio level 4)
  • gas_state (Gas state)
  • calibration_level_5 (Calibration ratio level 5)
  • o3_lamp_state (Ozone lamp state)
  • lamp_temperature (Lamp temperature)
  • noise_a (Electric noise in cell A)
  • diagnostic_voltage_mib_3p3 (Diagnostic +3.3 volts at measurement interface board)
  • o3_lamp_temperature (Ozone lamp temperature)
  • averaging_time (Instrument averaging time)
  • flow_a (Flow in cell A)
  • intensity_a (Intensity in cell A)
  • lamp_voltage_ozonizer (Lamp voltage ozonizer)
  • o3_background (Ozone background)
  • calibration_level_1 (Calibration ratio level 1)
  • range (Instrument range setting)
  • diagnostic_voltage_mb_15 (Diagnostic +15 volts at motherboard)
  • diagnostic_voltage_mb_5 (Diagnostic +5 volts at motherboard)
  • o3 (Ozone concentration)
  • diagnostic_voltage_mb_minus_3p3 (Diagnostic -3.3 volts at motherboard)
  • o3_pressure (Ozone pressure)
  • temperature_compensation_state (Temperature compensation state)
  • o3_bench_temperature (Bench temperature)
  • noise_b (Electric noise in cell B)
  • diagnostic_voltage_mib_minus_15 (Diagnostic -15 volts at measurement interface board)
  • diagnostic_voltage_mb_24 (Diagnostic +24 volts at motherboard)
  • calibration_level_3 (Calibration ratio level 3)
 
DQR ID: D160621.4    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-06-16 00:42:00 2016-06-16 13:45:00 Incorrect
Subject: ASI/AOS/M1 - Manifold blower not working
Affected Datastreams: asiaosccn2colbavgM1.b1
Description Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic.
Measurements
  • lat (North latitude)
  • base_time (Base time in Epoch)
  • lon (East longitude)
  • time_offset (Time offset from base_time)
  • averaging_time (Instrument averaging time)
  • alt (Altitude above mean sea level)
  • time (Time offset from midnight)
 
DQR ID: D170208.4    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-12-14 01:57:16 2016-12-14 09:51:51 Missing
Subject: ASI/OZONE/TRACEGAS/M1 - Instrument computer failure
Affected Datastreams: asiaosso2M1.a1
Description Unit three computer failed. No data recorded until computer rebuilt remotely.
Measurements
  • pressure_compensation_state (Pressure compensation state)
  • intensity_b (Intensity in cell B)
  • o3_coefficient (Instrument ozone coefficient)
  • lamp_level (Lamp level)
  • calibration_level_2 (Calibration ratio level 2)
  • diagnostic_voltage_mib_24 (Diagnostic +24 volts at measurement interface board)
  • flow_b (Flow in cell B)
  • lamp_voltage_bench (Lamp bench voltage)
  • diagnostic_voltage_mib_5 (Diagnostic +5 volts at measurement interface board)
  • alt (Altitude above mean sea level)
  • diagnostic_voltage_mb_3p3 (Diagnostic +3.3 volt at motherboard)
  • o3_flags (Ozone flag string)
  • diagnostic_voltage_mib_15 (Diagnostic +15 volts at measurement interface board)
  • calibration_level_4 (Calibration ratio level 4)
  • gas_state (Gas state)
  • calibration_level_5 (Calibration ratio level 5)
  • lat (North latitude)
  • base_time (Base time in Epoch)
  • o3_lamp_state (Ozone lamp state)
  • lamp_temperature (Lamp temperature)
  • noise_a (Electric noise in cell A)
  • intensity_a (Intensity in cell A)
  • lamp_voltage_ozonizer (Lamp voltage ozonizer)
  • averaging_time (Instrument averaging time)
  • diagnostic_voltage_mib_3p3 (Diagnostic +3.3 volts at measurement interface board)
  • flow_a (Flow in cell A)
  • o3_lamp_temperature (Ozone lamp temperature)
  • o3_background (Ozone background)
  • calibration_level_1 (Calibration ratio level 1)
  • range (Instrument range setting)
  • diagnostic_voltage_mb_15 (Diagnostic +15 volts at motherboard)
  • diagnostic_voltage_mb_5 (Diagnostic +5 volts at motherboard)
  • o3 (Ozone concentration)
  • diagnostic_voltage_mb_minus_3p3 (Diagnostic -3.3 volts at motherboard)
  • o3_pressure (Ozone pressure)
  • temperature_compensation_state (Temperature compensation state)
  • o3_bench_temperature (Bench temperature)
  • noise_b (Electric noise in cell B)
  • lon (East longitude)
  • time_offset (Time offset from base_time)
  • diagnostic_voltage_mib_minus_15 (Diagnostic -15 volts at measurement interface board)
  • time (Time offset from midnight)
  • diagnostic_voltage_mb_24 (Diagnostic +24 volts at motherboard)
  • calibration_level_3 (Calibration ratio level 3)
 
DQR ID: D170104.3    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-04-23 11:56:00 2016-11-02 17:00:59 Note
Subject: ASI/OZONE/M1 - Reprocessed and Corrected: asiaosozoneM1.a1 to asiaoso3M1.a1
Affected Datastreams: asiaoso3M1.a1
Description DATA REPROCESSED and CORRECTED: The aosozone.a1 data was renamed to aoso3.a1. In addition to the rename, we added data from raw data columns that were originally not included. These additional columns will be used in subsequent downstream processing. The data were archived February 2017. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-02 REPROCESS IDENTIFICATION NUMBER: RID 1526 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The aosozone.a1 data is being renamed to aoso3.a1. In addition to the rename, we are adding data from raw data columns that were missed. These additional columns will be used in subsequent downstream processing.
Measurements
  • lat (North latitude)
  • base_time (Base time in Epoch)
  • lon (East longitude)
  • time_offset (Time offset from base_time)
  • alt (Altitude above mean sea level)
  • time (Time offset from midnight)
 
DQR ID: D170208.3    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-11-21 20:46:16 2016-11-22 13:00:49 Incorrect
Subject: ASI/OZONE/TRACEGAS/M1 - Trace gas inlet blower failure
Affected Datastreams: asiaoso3M1.a1
Description MAOS C Trace gas inlet had the blower die. Though there are numbers, they should be viewed as incorrect. It also affects the PTR-MS.
Measurements
  • pressure_compensation_state (Pressure compensation state)
  • intensity_b (Intensity in cell B)
  • o3_coefficient (Instrument ozone coefficient)
  • lamp_level (Lamp level)
  • calibration_level_2 (Calibration ratio level 2)
  • flow_b (Flow in cell B)
  • diagnostic_voltage_mib_24 (Diagnostic +24 volts at measurement interface board)
  • lamp_voltage_bench (Lamp bench voltage)
  • diagnostic_voltage_mib_5 (Diagnostic +5 volts at measurement interface board)
  • diagnostic_voltage_mb_3p3 (Diagnostic +3.3 volt at motherboard)
  • o3_flags (Ozone flag string)
  • diagnostic_voltage_mib_15 (Diagnostic +15 volts at measurement interface board)
  • calibration_level_4 (Calibration ratio level 4)
  • gas_state (Gas state)
  • calibration_level_5 (Calibration ratio level 5)
  • lamp_temperature (Lamp temperature)
  • o3_lamp_state (Ozone lamp state)
  • noise_a (Electric noise in cell A)
  • diagnostic_voltage_mib_3p3 (Diagnostic +3.3 volts at measurement interface board)
  • lamp_voltage_ozonizer (Lamp voltage ozonizer)
  • averaging_time (Instrument averaging time)
  • intensity_a (Intensity in cell A)
  • flow_a (Flow in cell A)
  • o3_lamp_temperature (Ozone lamp temperature)
  • o3_background (Ozone background)
  • calibration_level_1 (Calibration ratio level 1)
  • range (Instrument range setting)
  • diagnostic_voltage_mb_15 (Diagnostic +15 volts at motherboard)
  • o3 (Ozone concentration)
  • diagnostic_voltage_mb_5 (Diagnostic +5 volts at motherboard)
  • diagnostic_voltage_mb_minus_3p3 (Diagnostic -3.3 volts at motherboard)
  • o3_pressure (Ozone pressure)
  • temperature_compensation_state (Temperature compensation state)
  • o3_bench_temperature (Bench temperature)
  • noise_b (Electric noise in cell B)
  • diagnostic_voltage_mib_minus_15 (Diagnostic -15 volts at measurement interface board)
  • diagnostic_voltage_mb_24 (Diagnostic +24 volts at motherboard)
  • calibration_level_3 (Calibration ratio level 3)