Data Quality Report for Order ID: 187061
User ID: endos1    Date completed: 2017-08-05
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D161005.3 SGP/TSI/C1 - Reprocess: Corrupt properties file sgptsiskycoverC1.b1 Incorrect
D170228.1 SGP/TSI/C1 - Reprocess: Global attribute units incorrect for cam_mir_dist sgptsiskycoverC1.b1 Incorrect
 
DQR ID: D161005.3    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2014-05-14 14:02:00 2016-09-23 22:01:59 Incorrect
Subject: SGP/TSI/C1 - Reprocess: Corrupt properties file
Affected Datastreams: sgptsiskycoverC1.b1
Description The YYYMMDDHHMMSS_tsi.properties files were corrupted when the computer operating system was upgraded, causing the ingest to assign incorrect values to attributes opaque_thresh, thin_thresh, cam_mir_dist, and center_x. A default properties file from 5/25/2004 was briefly used and then a properties file from 10/11/2013 was being concatenated to the current properties file before it was moved to the ingest directory, so that the last values read were from the earlier date.
Measurements
  • base_time (Base time in Epoch)
 
DQR ID: D170228.1    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2000-07-02 00:00:00 2017-01-20 15:59:59 Incorrect
Subject: SGP/TSI/C1 - Reprocess: Global attribute units incorrect for cam_mir_dist
Affected Datastreams: sgptsiskycoverC1.b1
Description The global attribute cam_mir_dist was converted incorrectly. The raw camera.mirror.distance, in inches, was multiplied by 10 and assigned a unit of mm.
Measurements
  • base_time (Base time in Epoch)