Data Quality Report for Order ID: 187101
User ID: eluke    Date completed: 2017-08-07
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D160610.6 AWR/KAZR/M1 - System Error awrkazrspeccmaskgecopolM1.a0 Incorrect
D160610.7 AWR/KAZR/M1 - Transmit Power Low awrkazrspeccmaskgecopolM1.a0 Suspect
 
DQR ID: D160610.6    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-05-30 16:25:00 2016-05-30 19:05:00 Incorrect
Subject: AWR/KAZR/M1 - System Error
Affected Datastreams: awrkazrspeccmaskgecopolM1.a0
Description The TWTA had an error and the radar was not transmitting during this period. Returns are just background noise.
Measurements
  • locator_mask (Locator mask for spectra)
  • spectra (Spectra)
 
DQR ID: D160610.7    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-05-30 17:00:00 2016-06-05 01:00:00 Suspect
Subject: AWR/KAZR/M1 - Transmit Power Low
Affected Datastreams: awrkazrspeccmaskgecopolM1.a0
Description The data look OK, but the transmit power was reported at ~25dB vs the normal 51dB. There was likely a sensor error with reading the transmitter output, but be cautious when using the data either way.
Measurements
  • locator_mask (Locator mask for spectra)
  • spectra (Spectra)