Data Quality Report for Order ID: 187105
User ID: zhangj1    Date completed: 2017-08-07
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D170316.1 ASI/CEIL/S1 - Reprocess: Incorrect averaging_inverval global attribute asiceilS1.b1 Note
 
DQR ID: D170316.1    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-05-24 17:30:00 2016-11-14 15:59:59 Note
Subject: ASI/CEIL/S1 - Reprocess: Incorrect averaging_inverval global attribute
Affected Datastreams: asiceilS1.b1
Description When the ceilometer software was changed from CT-View to CL-View, several firmware configuration changes were made to the model CT25K for consistency with the CL31 at other sites, including the data acquisition interval which was changed from 15 to 16 seconds. However, the global attribute "averaging_interval" in the ceil datastream was not changed.
Measurements
  • base_time (Base time in Epoch)