DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D140815.1 | MAO/AOS/S1/M1 - Instrument maintenance | maoaosacsmS1.a1 | Incorrect |
D140822.4 | MAO/ACSM/S1 - instrument maintenance | maoaosacsmS1.a1 | Missing |
D150921.2 | MAO/AOS/M1/S1 - Forest Fire | maoaosacsmS1.a1 | Note |
D150917.6 | MAO/ACSM/AETH/CPC/NEPHELOMETER/PSAP/SMPS/UHSAS/S1 Calibration of the Nephelometer | maoaosacsmS1.a1 | Incorrect |
D150414.3 | MAO/ACSM/S1 - Noisy signal | maoaosacsmS1.a1 | Suspect |
D150824.2 | MAO/ACSM/S1 - Incorrect and missing data | maoaosacsmS1.a1 | Incorrect |
D150702.6 | MAO/ACSM/S1 - Critical orifice cleaning | maoaosacsmS1.a1 | Missing |
D150619.7 | MAO/ACSM/AETH/CCN/CPC/HTDMA/NEPHELOMETER/PASS/PSAP/SMPS/SP2/UHSAS/S1 - Manifold blower stopped | maoaospsap3wS1.b1 | Incorrect |
D140805.2 | MAO/ACSM/S1 - data collection problem | maoaosacsmS1.a1 | Missing |
D140501.2 | MAO/ACSM/S1 - Data acquisition problem | maoaosacsmS1.a1 | Missing |
D150429.6 | MAO/AOS/S1 - Flow rate change | maoaosacsmS1.a1 | Note |
D151019.5 | MAO/ACSM/S1 - critical orifice clog and resulting noise issues | maoaosacsmS1.a1 | Incorrect |
D151008.2 | MAO/ACSM/S1 - Instrument noise | maoaosacsmS1.a1 | Suspect |
D140811.1 | MAO/ACSM/S1 - Instrument tuning and calibration | maoaosacsmS1.a1 | Missing |
D150922.2 | MAO/ACSM/S1 - Fluctuations in inlet pressure | maoaosacsmS1.a1 | Incorrect |
D150626.3 | MAO/ACSM/S1 - Instrument shutdown | maoaosacsmS1.a1 | Missing |
D141203.3 | MAO/ALL/S1 - Inlet down/off for cleaning and maintenance. | maoaosacsmS1.a1 | Incorrect |
D150916.2 | MAO/ACSM/S1 - instrument offline for calibration | maoaosacsmS1.a1 | Missing |
D151124.2 | MAO/ACSM/S1 - Data missing from controller failure | maoaosacsmS1.a1 | Missing |
D141009.5 | MAO/ACSM/S1 - Reprocess: Aerosol chemical speciation monitor errors in data | maoaosacsmS1.a1 | Incorrect |
D150619.6 | MAO/ACSM/S1 - Noise in data | maoaosacsmS1.b1 | Suspect |
D150223.1 | MAO/ACSM/S1 - Clogged critical orifice | maoaosacsmS1.a1 | Incorrect |
D140221.5 | MAO/ACSM/S1 - Missing data | maoaosacsmS1.a1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-08-09 | 16:00:00 | 2014-08-15 | 15:00:00 | Incorrect |
Subject: | MAO/AOS/S1/M1 - Instrument maintenance |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Instrument maintenance and cleaning inlets affecting data. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-08-21 | 15:19:00 | 2014-08-21 | 18:06:00 | Missing |
Subject: | MAO/ACSM/S1 - instrument maintenance |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Instrument was shut down for reconfiguration of UPS power. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-09-13 | 16:53:00 | 2015-09-13 | 18:10:00 | Note |
2015-10-04 | 00:00:00 | 2015-10-08 | 23:59:59 | Note |
2015-09-03 | 18:00:00 | 2015-09-03 | 23:59:00 | Note |
Subject: | MAO/AOS/M1/S1 - Forest Fire |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Smoke from forest burning can be seen on the TSI images rising and then subsequently moving over the site during this period. Many of the AOS instruments observed a sharp spike in data values. October 4-8 multiple fires in the area resulted in hazy conditions during this entire period. Most notable in the following instruments and variables ACSM - Total Organics AETH - Black Carbon Concentrations at all wavelengths CCN100 - Spike in number concentrations occurred a few hours after the smoke plume CLAP3W - Absorption Coefficient CPC - Similar to the CCN100, spike in particle concentration occurred a few hours later CPCF - Particle Concentration CPCU - Particle Concentration NEPHELOMETERS - Scattering/Backscattering Coeff PSAP3W - Absorption Coefficient SMPS - Concentration and Size Distribution UHSAS - Size Distribution |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-09-17 | 19:34:00 | 2015-09-17 | 20:35:00 | Incorrect |
Subject: | MAO/ACSM/AETH/CPC/NEPHELOMETER/PSAP/SMPS/UHSAS/S1 Calibration of the Nephelometer |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Inlet line for these instruments was used during Nephelometer calibration compromising the data. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-02-28 | 23:59:59 | 2014-07-20 | 00:00:00 | Suspect |
Subject: | MAO/ACSM/S1 - Noisy signal |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | At the end of February transient noise began to appear in the signal. This was not immediately apparent and was not detected until July 2015 when the noise suddenly stopped. It is apparent that the noise is periodic and probably electrical. Please see Document for the details. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-07-01 | 16:50:15 | 2015-07-08 | 01:01:52 | Incorrect |
Subject: | MAO/ACSM/S1 - Incorrect and missing data |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Data are incorrect or not available for the specified period. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-06-29 | 23:57:23 | 2015-07-01 | 16:50:14 | Missing |
Subject: | MAO/ACSM/S1 - Critical orifice cleaning |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Instrument out of service for critical orifice cleaning during the period specified. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-06-16 | 15:06:00 | 2015-06-17 | 13:21:59 | Incorrect |
Subject: | MAO/ACSM/AETH/CCN/CPC/HTDMA/NEPHELOMETER/PASS/PSAP/SMPS/SP2/UHSAS/S1 - Manifold blower stopped |
---|---|
Affected Datastreams: | maoaospsap3wS1.b1 |
Description | Manifold blower pulls air through the inlet to the instrument. Without the blower, the air source is indeterminant. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-07-31 | 12:00:00 | 2014-08-05 | 18:10:00 | Missing |
Subject: | MAO/ACSM/S1 - data collection problem |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | No data were collected for this period. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-04-29 | 12:00:00 | 2014-05-01 | 16:00:00 | Missing |
Subject: | MAO/ACSM/S1 - Data acquisition problem |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Data acquisition stopped because instrument parameters were incorrectly reset. Correction was made by the mentor and data acquisition was resumed. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-11-17 | 13:00:00 | 2015-04-20 | 18:45:59 | Note |
Subject: | MAO/AOS/S1 - Flow rate change |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | At begining of period, the bypass blower on MAOS A failed. This was replaced by the instrument operator, but operator was unable to reprogram the PID flow controller. The blower ran approximately 50% faster than designed flow rate. It is unlikely this significantly affected the flow dynamics or transmission efficiencies in the MAOS A sampling stack. The data is colored transparent, but the DQR is filed to alert the user. No specific degradation of signals were noted at this time. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-10-04 | 21:59:11 | 2015-10-13 | 20:02:04 | Incorrect |
Subject: | MAO/ACSM/S1 - critical orifice clog and resulting noise issues |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | The critical orifice clogged at the beginning of the specified period and there was a substantial drop in the inlet pressure. the orifice was cleaned and reinstalled, but there was significant instrument noise. A second cleaning and reinstallation solved the problem. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-09-18 | 03:07:54 | 2015-09-19 | 13:13:04 | Suspect |
2015-08-25 | 15:12:00 | 2015-08-29 | 18:03:59 | Suspect |
Subject: | MAO/ACSM/S1 - Instrument noise |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Data is noisy and should be considered suspect. Data is not useful for PMF calculations |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-08-09 | 19:17:05 | 2014-08-10 | 16:19:26 | Missing |
Subject: | MAO/ACSM/S1 - Instrument tuning and calibration |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | The instrument was off line for calibration and tuning during the period specified. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-08-25 | 21:50:55 | 2015-08-26 | 23:53:44 | Incorrect |
Subject: | MAO/ACSM/S1 - Fluctuations in inlet pressure |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Fluctuations in inlet pressure possibly as a result of power and/or pump problems. Data are suspect for the specified time period. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-06-24 | 18:04:05 | 2015-06-26 | 15:35:20 | Missing |
Subject: | MAO/ACSM/S1 - Instrument shutdown |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | The data acquisition program locked up on 2015-06-24 and no data were collected. Program was restarted 2015-06-26 15:06:39. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-08-10 | 13:25:45 | 2014-08-10 | 21:15:27 | Incorrect |
Subject: | MAO/ALL/S1 - Inlet down/off for cleaning and maintenance. |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Inlet stack of MAOS A lowered. All instruments on internal cabin air. No atmospheric validity on HTDMA, PSAP, f(RH) dry neph, wet neph, aethalometer, PASS3, SP2, ACSM, CCN-200, Impactor, CPC, UCPC, SMPS. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-09-15 | 14:54:12 | 2015-09-16 | 13:59:15 | Missing |
Subject: | MAO/ACSM/S1 - instrument offline for calibration |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Data are unavailable during the specified period |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-11-11 | 16:54:00 | 2015-12-01 | 23:59:59 | Missing |
Subject: | MAO/ACSM/S1 - Data missing from controller failure |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | The electronics controlling the quadrupole mass spectrometer have failed. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-02-09 | 00:00:00 | 2014-10-09 | 23:59:59 | Incorrect |
Subject: | MAO/ACSM/S1 - Reprocess: Aerosol chemical speciation monitor errors in data |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | The data will be reprocessed with the manufacturer's software to correct for collection efficiencies and instrument response. The data in the current form is incorrect and should not be used. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-02-09 | 00:00:00 | 2015-08-04 | 14:10:00 | Suspect |
Subject: | MAO/ACSM/S1 - Noise in data |
---|---|
Affected Datastreams: | maoaosacsmS1.b1 |
Description | There continue to be issues with noise in the ACSM in the MAOS A container located at the site near Manacapuru in Amazonas, Brazil. The noise is intermittent, ongoing, and persistent. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-02-03 | 23:00:00 | 2015-02-08 | 17:00:00 | Incorrect |
Subject: | MAO/ACSM/S1 - Clogged critical orifice |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Inlet pressure was low because of a clogged critical orifice. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-02-17 | 21:00:00 | 2014-02-18 | 19:30:00 | Missing |
2014-02-19 | 22:30:00 | 2014-02-20 | 22:00:00 | Missing |
2014-02-11 | 12:00:00 | 2014-02-12 | 14:00:00 | Missing |
Subject: | MAO/ACSM/S1 - Missing data |
---|---|
Affected Datastreams: | maoaosacsmS1.a1 |
Description | Data are missing and unrecoverable for the specified periods. |