DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D120913.3 | NSA/CPC/X1 - Global metadata error | nsaaoscpcX1.a1 | Note |
D150701.24 | NSA/CPC/X1 - Reprocess: reprocessing from raw data in preparation for not-yet-released b1 level | nsaaoscpcX1.a1 | Note |
D160927.4 | NSA/AOS/CLAP/CPC/NEPHELOMETER/X1 - Site power outage | nsaaospsap3wX1.b1 | Missing |
D160927.3 | NSA/AOS/CLAP/CPC/NEPHELOMETER/X1 - System maintenance | nsaaoscpcX1.a1 | Incorrect |
D160614.5 | NSA/AOS/AETH/CLAP/CPC/NEPHELOMETER/X1 - Data missing | nsaaosnephdry1mX1.b1 | Missing |
D160803.8 | NSA/AOS/AETH/CLAP/CPC/NEPHELOMETER/X1 - data collection error | nsaaosclap3w1mX1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2011-09-01 | 01:00:00 | 2012-07-06 | 12:00:00 | Note |
Subject: | NSA/CPC/X1 - Global metadata error |
---|---|
Affected Datastreams: | nsaaoscpcX1.a1 |
Description | The global metadata in the nsaaoscpcX1.a1 file incorrectly specifies the CPC model number as "3372". The correct model number is "3010". This is a metadata error only. The data are not affected. A low priority reprocessing task has been submitted. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2011-09-01 | 00:00:00 | 2015-10-01 | 23:59:59 | Note |
Subject: | NSA/CPC/X1 - Reprocess: reprocessing from raw data in preparation for not-yet-released b1 level |
---|---|
Affected Datastreams: | nsaaoscpcX1.a1 |
Description | End to end reprocessing to generate new a1 datastreams is needed so that when the harmonization software is released to production, we will be ready with the a1 datastreams that are input to the subsequent processing software. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-08-21 | 20:52:00 | 2016-08-22 | 13:40:00 | Missing |
Subject: | NSA/AOS/CLAP/CPC/NEPHELOMETER/X1 - Site power outage |
---|---|
Affected Datastreams: | nsaaospsap3wX1.b1 |
Description | Site power outage |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-06-16 | 18:43:00 | 2016-08-20 | 18:47:00 | Incorrect |
Subject: | NSA/AOS/CLAP/CPC/NEPHELOMETER/X1 - System maintenance |
---|---|
Affected Datastreams: | nsaaoscpcX1.a1 |
Description | Data is either missing or instrument was being calibrated |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-06-10 | 20:07:00 | 2016-06-13 | 15:33:00 | Missing |
Subject: | NSA/AOS/AETH/CLAP/CPC/NEPHELOMETER/X1 - Data missing |
---|---|
Affected Datastreams: | nsaaosnephdry1mX1.b1 |
Description | Communication failure between data acquisition and instruments. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-08-01 | 22:00:00 | 2016-08-02 | 16:50:00 | Missing |
Subject: | NSA/AOS/AETH/CLAP/CPC/NEPHELOMETER/X1 - data collection error |
---|---|
Affected Datastreams: | nsaaosclap3w1mX1.b1 |
Description | Hardware failure of data acquisition laptop. Laptop replaced. |
Measurements |
|