Data Quality Report for Order ID: 187220
User ID: chenm2    Date completed: 2017-08-11
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D160524.3 SGP/OZONE/S01 - Filter not changed sgpaoso3S01.a1 Suspect
D160610.13 SGP/S01 - Tower inlet not operational sgpaosso2S01.a1 Incorrect
D161117.4 SGP/TRACEGAS/S01 - Instrument sgpaospsaprS01.a0 Missing
D160824.8 SGP/OZONE/S01 - Reprocessed and Corrected: sgpaosozoneS01.a1 to sgpaoso3S01.a1 sgpaoso3S01.a1 Note
 
DQR ID: D160524.3    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-03-25 00:00:00 2016-04-19 17:44:00 Suspect
Subject: SGP/OZONE/S01 - Filter not changed
Affected Datastreams: sgpaoso3S01.a1
Description Filter not changed. Pressure low and response to calibrations increased. The time before the filter change is considered suspect.
Measurements
  • ozone_bench_temp (Ozone bench temperature)
  • ozone (Ozone concentration)
  • ozone_pressure (Ozone pressure)
 
DQR ID: D160610.13    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-05-05 14:23:50 2016-05-05 17:40:20 Incorrect
Subject: SGP/S01 - Tower inlet not operational
Affected Datastreams: sgpaosso2S01.a1
Description Inlet not operational in order to change met sensor.
Measurements
  • diagnostic_voltage_mib_15 (Diagnostic +15 volts at measurement interface board)
  • pressure_compensation_state (Pressure compensation state)
  • diagnostic_voltage_mb_minus_3p3 (Diagnostic -3.3 volts at motherboard)
  • temperature_compensation_state (Temperature compensation state)
  • ozone_bench_temp (Ozone bench temperature)
  • range (Instrument range setting)
  • diagnostic_voltage_mib_minus_15 (Diagnostic -15 volts at measurement interface board)
  • diagnostic_voltage_mib_3p3 (Diagnostic +3.3 volts at measurement interface board)
  • diagnostic_voltage_mb_3p3 (Diagnostic +3.3 volt at motherboard)
  • averaging_time (Instrument averaging time)
  • diagnostic_voltage_mib_5 (Diagnostic +5 volts at measurement interface board)
  • ozone_pressure (Ozone pressure)
  • diagnostic_voltage_mb_24 (Diagnostic +24 volts at motherboard)
  • diagnostic_voltage_mb_5 (Diagnostic +5 volts at motherboard)
  • diagnostic_voltage_mb_15 (Diagnostic +15 volts at motherboard)
  • ozone (Ozone concentration)
  • diagnostic_voltage_mib_24 (Diagnostic +24 volts at measurement interface board)
 
DQR ID: D161117.4    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-03-27 08:34:00 2016-03-28 20:00:00 Missing
2016-04-17 10:21:27 2016-04-18 14:15:14 Missing
Subject: SGP/TRACEGAS/S01 - Instrument
Affected Datastreams: sgpaospsaprS01.a0
Description Improper start up following a power failure. Computers came back up but instruments did not restart (by design for Oliktok). Daily checklists not compiled this period (weekend).
Measurements
  • range (Instrument range setting)
  • lat (North latitude)
  • averaging_time (Instrument averaging time)
  • lon (East longitude)
  • diagnostic_voltage_mb_24 (Diagnostic +24 volts at motherboard)
  • diagnostic_voltage_mib_24 (Diagnostic +24 volts at measurement interface board)
  • base_time (Base time in Epoch)
  • pressure_compensation_state (Pressure compensation state)
  • ozone_bench_temp (Ozone bench temperature)
  • diagnostic_voltage_mib_3p3 (Diagnostic +3.3 volts at measurement interface board)
  • alt (Altitude above mean sea level)
  • diagnostic_voltage_mb_3p3 (Diagnostic +3.3 volt at motherboard)
  • time_offset (Time offset from base_time)
  • diagnostic_voltage_mb_5 (Diagnostic +5 volts at motherboard)
  • time (Time offset from midnight)
  • diagnostic_voltage_mib_15 (Diagnostic +15 volts at measurement interface board)
  • diagnostic_voltage_mb_minus_3p3 (Diagnostic -3.3 volts at motherboard)
  • temperature_compensation_state (Temperature compensation state)
  • diagnostic_voltage_mib_minus_15 (Diagnostic -15 volts at measurement interface board)
  • diagnostic_voltage_mib_5 (Diagnostic +5 volts at measurement interface board)
  • diagnostic_voltage_mb_15 (Diagnostic +15 volts at motherboard)
  • ozone (Ozone concentration)
  • ozone_pressure (Ozone pressure)
 
DQR ID: D160824.8    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-02-02 00:00:00 2016-06-10 23:59:00 Note
Subject: SGP/OZONE/S01 - Reprocessed and Corrected: sgpaosozoneS01.a1 to sgpaoso3S01.a1
Affected Datastreams: sgpaoso3S01.a1
Description DATA REPROCESSED and CORRECTED: The aosozone.a1 data was renamed to aoso3.a1. In addition to the rename, we added data from raw data columns that were originally not included. These additional columns will be used in subsequent downstream processing. The data were archived February 2017. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-02 REPROCESS IDENTIFICATION NUMBER: RID 1496 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The sgpaosozoneA01.a1 data is being renamed to aoso3.a1. In addition to the rename, we are adding data from raw data columns that were missed. These additional columns will be used in subsequent downstream processing.
Measurements
  • base_time (Base time in Epoch)
  • ozone_bench_temp (Ozone bench temperature)
  • alt (Altitude above mean sea level)
  • ozone (Ozone concentration)
  • lat (North latitude)
  • lon (East longitude)
  • time_offset (Time offset from base_time)
  • time (Time offset from midnight)
  • ozone_pressure (Ozone pressure)