Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2011-01-18 | 00:00:00 | 2014-02-07 | 23:59:00 | Incorrect |
Subject: | NSA/KAZR/C1 - Reprocess: Incorrect calibration coefficients |
---|---|
Affected Datastreams: | nsakazrspeccmaskgexpolC1.a0 |
Description | The wrong calibrations were used due to a problem with the software. Mentor discovered the reflectivity between long and short pulse did not agree with each other after analyzing data. Data will be reprocessed. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2012-12-14 | 16:50:00 | 2013-03-17 | 18:35:00 | Suspect |
Subject: | NSA/KAZR/C1 - Shift in Background Reflectivity and Signal-to-Noise Ratio |
---|---|
Affected Datastreams: | nsakazrspeccmaskgexpolC1.a0 |
Description | The KAZR experienced an abrupt shift in the background reflectivity on 12/14/2012, followed by a shift in the background Signal-to-Noise Ratio on 12/16/2012. The background levels then revert back to "normal" levels on 3/17/2013. Cause is currently unknown. |