Data Quality Report for Order ID: 187240
User ID: eluke    Date completed: 2017-08-14
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D131120.3 NSA/KAZR/C1 - Reprocess: Incorrect calibration coefficients nsakazrspeccmaskgexpolC1.a0 Incorrect
 
DQR ID: D131120.3    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2011-01-18 00:00:00 2014-02-07 23:59:00 Incorrect
Subject: NSA/KAZR/C1 - Reprocess: Incorrect calibration coefficients
Affected Datastreams: nsakazrspeccmaskgexpolC1.a0
Description The wrong calibrations were used due to a problem with the software. Mentor discovered the reflectivity between long and short pulse did not agree with each other after analyzing data. Data will be reprocessed.
Measurements
  • reflectivity (Reflectivity)