Data Quality Report for Order ID: 187251
User ID: wehnerb1    Date completed: 2017-08-15
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D170726.7 ENA/CO-ANALYZER/C1 - Reprocess: Valve position fields incorrect enaaoscoC1.b1 Incorrect
 
DQR ID: D170726.7    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2017-06-13 18:00:00 2017-07-26 23:59:59 Incorrect
Subject: ENA/CO-ANALYZER/C1 - Reprocess: Valve position fields incorrect
Affected Datastreams: enaaoscoC1.b1
Description A bug in the a1- and b1-level AOS CO ingest processes caused the valve position fields to be incorrect. This had the trickle-down effect of flagging the primary fields as incorrect via embedded QC in the b1-level product. Reprocessing is being requested now that this ingest bug has been corrected.
Measurements
  • qc_h2o (Quality check results on field: Water vapor mixing ratio calculated with nominal sensitivity correction)
  • qc_n2o (Quality check results on field: Nitrous oxide (N2O) mixing ratio calculated with nominal sensitivity correction)
  • qc_n2o_dry (Quality check results on field: Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and calculated with nominal sensitivity correction)
  • valve_position_MFC_2 (Valve position mass flow controller 2)
  • valve_position_MFC_1 (Valve position for mass flow controller 1)
  • qc_co (Quality check results on field: Carbon monoxide (CO) mixing ratio calculated with nominal sensitivity correction)
  • qc_co_dry (Quality check results on field: Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and calculated with nominal sensitivity correction)