DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D170110.2 | OLI/AOS/M1 - Severe storm | oliaosnephwetM1.a1 | Note |
D161209.2 | OLI/TRACEGAS/M1 - Weekly displacement calibration of CO did not occur | oliaoscoM1.a1 | Note |
D170303.14 | OLI/TRACEGAS/M1 - Multiple dropouts in data | oliaoscoM1.a1 | Missing |
D170320.6 | OLI/ALL/M1 - Power Outage | oliaosnephwet1mM1.b1 | Note |
D170405.5 | OLI/TRACEGAS/M1 - Instrument communication issue | oliaoscoM1.a1 | Missing |
D170405.4 | OLI/TRACEGAS/M1 - Fire in generator | oliaoscoM1.a1 | Missing |
D170404.2 | OLI/TRACEGAS/M1 - Power failure at site | oliaoscoM1.a1 | Missing |
D170511.3 | OLI/TRACEGAS/M1 - Data output changed to 0.5 Hz | oliaoscoM1.a1 | Note |
D170405.3 | OLI/TRACEGAS/M1 - Power failure and poor restart | oliaoscoM1.a1 | Missing |
D170726.8 | OLI/CO-ANALYZER/M1 - Reprocess: Valve position fields incorrect | oliaoscoM1.b1 | Incorrect |
D170420.8 | OLI/AOS/M1 - Serial port server malfunction | oliaospsap3wM1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-01-05 | 14:00:00 | 2017-01-09 | 20:15:59 | Note |
Subject: | OLI/AOS/M1 - Severe storm |
---|---|
Affected Datastreams: | oliaosnephwetM1.a1 |
Description | The site lost power during a severe winter storm. Operators on site were not able to get out and restart the generators to bring AOS instrument systems back up until 1/9/2017. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-10-27 | 00:00:00 | 2016-12-02 | 00:00:59 | Note |
Subject: | OLI/TRACEGAS/M1 - Weekly displacement calibration of CO did not occur |
---|---|
Affected Datastreams: | oliaoscoM1.a1 |
Description | Some (minor) uncertainty in the calibration over this period. However, Calibrations before and after the event appear very stable. There are also daily standard additions between the weekly displacement calibrations. These daily checks are quite consistent. So this DQR is merely to notify of the condition. Data can be used with confidence in the judgment of the Mentor. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-02-20 | 16:00:00 | 2017-02-20 | 17:45:02 | Missing |
2017-02-20 | 19:54:22 | 2017-02-22 | 20:02:37 | Missing |
2017-02-05 | 00:00:00 | 2017-02-05 | 02:26:33 | Missing |
2017-02-01 | 17:00:00 | 2017-02-01 | 18:22:43 | Missing |
2017-02-11 | 23:00:00 | 2017-02-12 | 17:19:57 | Missing |
2017-02-20 | 19:54:00 | 2017-02-22 | 20:00:00 | Missing |
2017-02-11 | 17:55:35 | 2017-02-11 | 19:02:01 | Missing |
2017-02-20 | 03:00:00 | 2017-02-20 | 04:42:54 | Missing |
Subject: | OLI/TRACEGAS/M1 - Multiple dropouts in data |
---|---|
Affected Datastreams: | oliaoscoM1.a1 |
Description | Multiple dropouts in data. These appear associated with an ongoing problem with the serial server and power interruptions at the site. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-03-06 | 18:47:01 | 2017-03-09 | 23:35:58 | Note |
Subject: | OLI/ALL/M1 - Power Outage |
---|---|
Affected Datastreams: | oliaosnephwet1mM1.b1 |
Description | OLI suffered from a prolonged power outage. The instruments were brought back up over the course of 3/9 so the end date will be different for each instrument. Any problems with the instruments after being brought back online will be covered by separate DQRs. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-04-12 | 07:00:00 | 2017-04-13 | 15:25:41 | Missing |
2017-03-31 | 15:00:00 | 2017-04-10 | 23:06:01 | Missing |
2017-04-11 | 16:00:00 | 2017-04-11 | 23:59:59 | Missing |
2017-03-29 | 16:00:00 | 2017-03-30 | 14:46:43 | Missing |
Subject: | OLI/TRACEGAS/M1 - Instrument communication issue |
---|---|
Affected Datastreams: | oliaoscoM1.a1 |
Description | Communication issue with instrument. Instrument operating but serial switch not working |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-03-06 | 21:00:00 | 2017-03-24 | 13:31:03 | Missing |
Subject: | OLI/TRACEGAS/M1 - Fire in generator |
---|---|
Affected Datastreams: | oliaoscoM1.a1 |
Description | Instruments not running due to power outage and freezing conditions. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-01-05 | 13:35:43 | 2017-01-09 | 20:17:35 | Missing |
Subject: | OLI/TRACEGAS/M1 - Power failure at site |
---|---|
Affected Datastreams: | oliaoscoM1.a1 |
Description | Power failure at site has resulted in no data. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-04-13 | 15:25:42 | 2017-05-10 | 21:57:41 | Note |
Subject: | OLI/TRACEGAS/M1 - Data output changed to 0.5 Hz |
---|---|
Affected Datastreams: | oliaoscoM1.a1 |
Description | In an effort to diagnoses data dropouts, the instrument output rate was changed from 1 Hz to 0.5 Hz. Since the instrument averages over the reported time and the dynamic time response is ~2-5 s, there is no appreciable loss in S/N or time response. Once tests are completed on the virtual machine and the LGR instrument, the reporting speed will be returned to 1 Hz. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-03-04 | 09:00:00 | 2017-03-05 | 16:44:49 | Missing |
Subject: | OLI/TRACEGAS/M1 - Power failure and poor restart |
---|---|
Affected Datastreams: | oliaoscoM1.a1 |
Description | Generator failure and tests. CO was not brought up immediately. Computers were on and writing missing values. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-06-13 | 18:00:00 | 2017-07-26 | 23:59:59 | Incorrect |
Subject: | OLI/CO-ANALYZER/M1 - Reprocess: Valve position fields incorrect |
---|---|
Affected Datastreams: | oliaoscoM1.b1 |
Description | A bug in the a1- and b1-level AOS CO ingest processes caused the valve position fields to be incorrect. This had the trickle-down effect of flagging the primary fields as incorrect via embedded QC in the b1-level product. Reprocessing is being requested now that this ingest bug has been corrected. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-04-18 | 11:40:00 | 2017-04-19 | 00:00:00 | Missing |
2017-04-12 | 07:30:00 | 2017-04-13 | 15:30:00 | Missing |
2017-03-31 | 15:00:00 | 2017-04-10 | 23:00:00 | Missing |
2017-03-29 | 16:20:00 | 2017-03-30 | 14:40:00 | Missing |
Subject: | OLI/AOS/M1 - Serial port server malfunction |
---|---|
Affected Datastreams: | oliaospsap3wM1.b1 |
Description | Serial port server malfunctioning. Data transmission from the instrument intermittent. |
Measurements |
|