Data Quality Report for Order ID: 187487
User ID: chenm2    Date completed: 2017-08-30
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D160524.3 SGP/OZONE/S01 - Filter not changed sgpaoso3S01.a1 Suspect
D160824.8 SGP/OZONE/S01 - Reprocessed and Corrected: sgpaosozoneS01.a1 to sgpaoso3S01.a1 sgpaoso3S01.a1 Note
D160610.13 SGP/S01 - Tower inlet not operational sgpaosso2S01.a1 Incorrect
D161117.4 SGP/TRACEGAS/S01 - Instrument sgpaospsaprS01.a0 Missing
 
DQR ID: D160524.3    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-03-25 00:00:00 2016-04-19 17:44:00 Suspect
Subject: SGP/OZONE/S01 - Filter not changed
Affected Datastreams: sgpaoso3S01.a1
Description Filter not changed. Pressure low and response to calibrations increased. The time before the filter change is considered suspect.
Measurements
  • ozone (Ozone concentration)
  • ozone_bench_temp (Ozone bench temperature)
  • ozone_pressure (Ozone pressure)
 
DQR ID: D160824.8    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-02-02 00:00:00 2016-06-10 23:59:00 Note
Subject: SGP/OZONE/S01 - Reprocessed and Corrected: sgpaosozoneS01.a1 to sgpaoso3S01.a1
Affected Datastreams: sgpaoso3S01.a1
Description DATA REPROCESSED and CORRECTED: The aosozone.a1 data was renamed to aoso3.a1. In addition to the rename, we added data from raw data columns that were originally not included. These additional columns will be used in subsequent downstream processing. The data were archived February 2017. Please retrieve the reprocessed data for use in your analysis. ORIGINAL DQR CATEGORY: Transparent (Does not affect quality) REPROCESSED and CORRECTED DATA ARCHIVED:2017-02 REPROCESS IDENTIFICATION NUMBER: RID 1496 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- The sgpaosozoneA01.a1 data is being renamed to aoso3.a1. In addition to the rename, we are adding data from raw data columns that were missed. These additional columns will be used in subsequent downstream processing.
Measurements
  • lon (East longitude)
  • time_offset (Time offset from base_time)
  • ozone (Ozone concentration)
  • base_time (Base time in Epoch)
  • ozone_bench_temp (Ozone bench temperature)
  • ozone_pressure (Ozone pressure)
  • lat (North latitude)
  • alt (Altitude above mean sea level)
  • time (Time offset from midnight)
 
DQR ID: D160610.13    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-05-05 14:23:50 2016-05-05 17:40:20 Incorrect
Subject: SGP/S01 - Tower inlet not operational
Affected Datastreams: sgpaosso2S01.a1
Description Inlet not operational in order to change met sensor.
Measurements
  • diagnostic_voltage_mb_5 (Diagnostic +5 volts at motherboard)
  • diagnostic_voltage_mb_15 (Diagnostic +15 volts at motherboard)
  • diagnostic_voltage_mb_minus_3p3 (Diagnostic -3.3 volts at motherboard)
  • averaging_time (Instrument averaging time)
  • pressure_compensation_state (Pressure compensation state)
  • diagnostic_voltage_mib_3p3 (Diagnostic +3.3 volts at measurement interface board)
  • range (Instrument range setting)
  • temperature_compensation_state (Temperature compensation state)
  • ozone_bench_temp (Ozone bench temperature)
  • diagnostic_voltage_mib_minus_15 (Diagnostic -15 volts at measurement interface board)
  • diagnostic_voltage_mb_3p3 (Diagnostic +3.3 volt at motherboard)
  • ozone (Ozone concentration)
  • diagnostic_voltage_mb_24 (Diagnostic +24 volts at motherboard)
  • ozone_pressure (Ozone pressure)
  • diagnostic_voltage_mib_24 (Diagnostic +24 volts at measurement interface board)
  • diagnostic_voltage_mib_15 (Diagnostic +15 volts at measurement interface board)
  • diagnostic_voltage_mib_5 (Diagnostic +5 volts at measurement interface board)
 
DQR ID: D161117.4    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2016-03-27 08:34:00 2016-03-28 20:00:00 Missing
2016-04-17 10:21:27 2016-04-18 14:15:14 Missing
Subject: SGP/TRACEGAS/S01 - Instrument
Affected Datastreams: sgpaospsaprS01.a0
Description Improper start up following a power failure. Computers came back up but instruments did not restart (by design for Oliktok). Daily checklists not compiled this period (weekend).
Measurements
  • diagnostic_voltage_mb_15 (Diagnostic +15 volts at motherboard)
  • lon (East longitude)
  • diagnostic_voltage_mb_3p3 (Diagnostic +3.3 volt at motherboard)
  • ozone_pressure (Ozone pressure)
  • diagnostic_voltage_mib_24 (Diagnostic +24 volts at measurement interface board)
  • time (Time offset from midnight)
  • averaging_time (Instrument averaging time)
  • diagnostic_voltage_mb_minus_3p3 (Diagnostic -3.3 volts at motherboard)
  • temperature_compensation_state (Temperature compensation state)
  • range (Instrument range setting)
  • base_time (Base time in Epoch)
  • ozone (Ozone concentration)
  • diagnostic_voltage_mib_minus_15 (Diagnostic -15 volts at measurement interface board)
  • diagnostic_voltage_mb_24 (Diagnostic +24 volts at motherboard)
  • diagnostic_voltage_mib_15 (Diagnostic +15 volts at measurement interface board)
  • time_offset (Time offset from base_time)
  • ozone_bench_temp (Ozone bench temperature)
  • lat (North latitude)
  • diagnostic_voltage_mb_5 (Diagnostic +5 volts at motherboard)
  • pressure_compensation_state (Pressure compensation state)
  • diagnostic_voltage_mib_3p3 (Diagnostic +3.3 volts at measurement interface board)
  • alt (Altitude above mean sea level)
  • diagnostic_voltage_mib_5 (Diagnostic +5 volts at measurement interface board)